Fluke i200 Manual


Læs nedenfor 📖 manual på dansk for Fluke i200 (2 sider) i kategorien Ikke kategoriseret. Denne guide var nyttig for 12 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/2
®
i200/i200s
AC Current Clamp
Instruction Sheet
Introducing the i200/i200s
The i200 is a single range 200A clamp-on AC Current Clamp with
current output via safety shrouded banana plugs.
The i200s is a dual range 20A and 200A clamp-on AC Current
Clamp with voltage output via a safety insulated BNC connector.
A dual banana to BNC adapter is supplied to allow the i200s to be
connected to multimeters with banana input.
Unpacking
The following items should be included in your Current Clamp box:
Current Clamp
Dual Banana to BNC Adapter model PM9081 (only with i200s)
Instruction Sheet (this paper)
Check the contents of the shipping box for completeness. If
something in the box has been damaged or missing, contact your
distributor or the nearest FLUKE sales or service office immediately.
Safety Information
Read First: Safety Information.
To ensure safe operation and service of the
current clamp, follow these instructions:
Read the operating instructions before use and follow all safety
instructions.
Use the Current Clamp only as specified in the operating
instructions, otherwise the clamp’s safety features may not
protect you.
Adhere to local and national safety codes. Individual protective
equipment must be used to prevent shock and arc blast injury
where hazardous live conductors are exposed.
Do not hold the Current Clamp anywhere beyond the tactile
barrier, see Figure 1.
Before each use, inspect the Current Clamp. Look for cracks or
missing portions of the clamp housing or output cable
insulation. Also look for loose or weakened components. Pay
particular attention to the insulation surrounding the jaws.
Check the magnetic mating surfaces of the clamp jaws; these
should be free of dust, dirt, rust and other foreign matter.
Never use the clamp on a circuit with voltages higher than 600
V CAT III.
CAT III equipment is designed to protect against transients
in equipment in fixed equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
Use extreme caution when working around bare conductors or
bus bars. Contact with the conductor could result in electric
shock.
Use caution when working with voltages above 60 V dc, 30 V ac
rms or 42 V ac peak. Such voltages pose a shock hazard.
Figure 1. Safely holding the Current Clamp
Symbols
May be used on HAZARDOUS LIVE
conductors.
Product is protected by double insulation.
Risk of Danger. Important information. See
Instruction Sheet.
Risk of Electric Shock.
Conforms to relevant European standards.
Earth ground.
Specifications
SAFETY
Input jaws & output floating
voltage to ground
Complies with American
industry standards
UL61010B-1 & UL61010B-2-
032 and European standards
EN/IEC 61010-1 2nd Edition
& EN/IEC 61010-02-032 for
600V CAT III, pollution
degree 2.
EMC Complies with standards
EN/IEC 50081-1 &
EN/IEC 50082-2
ELECTRICAL SPECIFICATIONS
All Electrical Specifications are valid at the following reference
conditions:
Ambient temperature 23±3°C (73±3°F)
Relative Humidity 20 to 75%
Frequency 48 to 65 Hz
Continuous external field < 40 A/m
Load impedance i200: 0.2
…15
i200s: >1 M
// 100 pF
The current may not contain any DC component
No influence from adjacent currents
The conductor must be centered within the jaw aperture
20A Range (i200s only)
Measuring range 0.1 to 24A
Maximum current 24A
Crest factor * < 3
Maximum non-destructive
current
200A (Frequency 1 kHz and
crest factor < 3)
Output signal 100 mV/A
Output impedance ≤ Ω 20 @ 1 kHz
Basic accuracy
48 Hz to 65 Hz 2% + 0.5A
Additional error:
40 Hz to 48 Hz and
65 Hz to 1 kHz
+ < 10%
1 kHz to 10 kHz + < 15%
Phase shift Unspecified
200A Range i200 i200s
Measuring range 0.5 to 240A 0.5 to 240A
Maximum current 240A 240A
Crest factor * < 3 < 3
Maximum non-destructive
current
@ Frequency 1 kHz and
crest factor < 3
Continuous 200A
10 min ON
/30 min OFF
240A
Output signal 1 mA/A 10 mV/A
Output impedance - ≤ Ω 10 @ 1 kHz
Basic accuracy
48 Hz to 65 Hz
0.5A to 10A ≤ ≤ 3% + 0.5A 3.5% + 0.5A
10A to 40A ≤ ≤ 2.5% + 0.5A 3% + 0.5A
40A to 100A ≤ ≤ 2% + 0.5A 2.5% + 0.5A
100A to 240A ≤ ≤ 1% + 0.5A 1.5% + 0.5A
Additional error:
40 Hz to 48 Hz and
65 Hz to 1 kHz
+ < 3%
+ < 3%
1 kHz to 10 kHz + < 12% + < 12%
Phase shift
0.5A to 10A Unspecified Unspecified
10A to 40A ≤ ≤ 5 ° 6 °
40A to 100A ≤ ≤ 3 ° 4 °
100A to 240A ≤ ≤ 2.5 ° 3 °
All ranges i200 i200s
Load on output 0.2...15 >1 M // < 100 pF
Load Influence Current: < 1%
Phase: < 1°
-
-
Bandwidth -1.5 dB
-3dB
40 Hz to 10 kHz
40 kHz
40 Hz to 10 kHz
40 kHz
Additional errors
With temperature 0.15 % / 10 K
With position of
conductor in the
clamp aperture
0.5 % @ 50 Hz
With adjacent
conductors
15 mA / A @ 50 Hz
* This is the maximum permissible ratio between the peak value
of a superimposed transient and the ac rms value.
GENERAL
Clamp Dimensions 135 x 50 x 30 mm
(5.3 x 2 x 1.2 in)
Protection index IP40
Jaw Opening 21 mm (0.82 in)
Height of open Jaws 69 mm (2.7 in)
Maximum conductor size 20 mm (0.8 in) or
busbar 20 x 5 mm (0.8 x 0.2 in)
Weight 180 g (6.4 oz)
Cable length i200
i200s
1.5 m (59 in)
2m (79 in)
Temperature
Operating
Non-operating
-10 to +55°C (+14 to +131°F)
-40 to + 70°C (-40 to +158°F)
Relative Humidity
Operating
85%, up to +30°C (+86°F)
75%, up to +55°C (+131°F)
Altitude
Operating
Non-operating
to 2000 m (6500 ft)
to 12000 m (40000 ft)
EMC EN/IEC 50081-1
EN/IEC 50082-2 (3V/m,
2.74V/yd)
Instrument Compatibility
The i200s is compatible with any Fluke ScopeMeter test tool, Power
Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage
measurement device that has the following features:
BNC input connector. The Dual Banana to BNC Adapter
included in the package, can be used to connect to standard
inputs on multimeters. For the 120 series ScopeMeters, use
the BB120 Shielded Banana to BNC Adapter.
Input accuracy of 2% or better to take full advantage of the
accuracy of the Current Clamp.
• Ω Input impedance of greater than or equal to 1 M , and for full
bandwidth and accuracy, a maximum input capacity of 100 pF.
A pass- band of more than four times the frequency of the
waveform to be measured.
The i200 is compatible with any Fluke Multimeter or any other
current measurement device that has the following features:
Banana inputs.
Input accuracy of 2% or better to take full advantage of the
accuracy of the Current Clamp.
Ω Ω Input impedance of 0.2 …15
A pass-band of more than four times the frequency of the
waveform to be measured.
4822 872 00934
July 1998, Rev. 5, 08/05
© 1998 - 2005 Fluke Corporation.
All rights reserved.
Printed in France
All product names are trademarks of
their respective companies.
Tactile barrier


Produkt Specifikationer

Mærke: Fluke
Kategori: Ikke kategoriseret
Model: i200

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke i200 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Fluke Manualer

Fluke

Fluke 438-II Manual

8 Oktober 2024
Fluke

Fluke 1670 Manual

8 Oktober 2024
Fluke

Fluke ii905 Manual

8 Oktober 2024
Fluke

Fluke ii915 Manual

8 Oktober 2024
Fluke

Fluke ii500 Manual

7 Oktober 2024
Fluke

Fluke T5-1000 Manual

4 Oktober 2024
Fluke

Fluke 700HTH Manual

29 September 2024
Fluke

Fluke 6200-2 Manual

22 September 2024
Fluke

Fluke Ti27 Manual

22 September 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer

Wagner

Wagner 2K Smart Manual

4 November 2024
Moen

Moen EB500-E Manual

3 November 2024
Moen

Moen Aspen 5824CH Manual

3 November 2024
Luxman

Luxman MQ-88uSE Manual

3 November 2024
Luxman

Luxman CL-38uSE Manual

3 November 2024
Luxman

Luxman CL-1000 Manual

3 November 2024
Moen

Moen Aspen 5802CH Manual

3 November 2024
Luxman

Luxman SQ-38u Manual

3 November 2024
Moen

Moen Aspen 5818CH Manual

3 November 2024