Installation Manual
Type
Medical Switching Power Supply
(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)
Installation
(1) Enclosed Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Enclosed units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Enclosed units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
(2) Modular Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Modular units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Modular units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.


Produkt Specifikationer

Mærke: Mean Well
Kategori: Ikke kategoriseret
Model: RPS-160-48

Har du brug for hjælp?

Hvis du har brug for hjælp til Mean Well RPS-160-48 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Mean Well Manualer

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer

Sonifex

Sonifex PS-AMP Manual

14 Januar 2025
Powerfix

Powerfix IAN 45827 Manual

13 Januar 2025
Faber

Faber Pisa 60 X Manual

13 Januar 2025
Hasbro

Hasbro A8752 Manual

13 Januar 2025
IFM

IFM EVC231 Manual

13 Januar 2025
Edimax

Edimax ES-5800G V2 Manual

13 Januar 2025
IFM

IFM PN2598 Manual

13 Januar 2025