IHealth BPM1 - Clear Manual

IHealth BPM1 - Clear

Læs nedenfor 📖 manual på dansk for IHealth BPM1 - Clear (8 sider) i kategorien . Denne guide var nyttig for 11 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/8
Wireless Blood Pressure Monitor (BPM1)
OWNER’S MANUAL
Monitor de tensão arterial sem fios (BPM1)
MANUAL DO PROPRIETÁRIO
MANUELE DEL PROPRIETARIO
dello sfigmomanometro wireless (BPM1)
Vernetztes-Blutdruckmessgerät (BPM1)
BENUTZERHANDBUCH
MANUAL DEL USUARIO
del tensiómetro inalámbrico (BPM1)
Draadloze bloeddrukmeter (BPM1)
GEBRUIKERSHANDLEIDING
Ασύρματο πιεσόμετρο (BPM1)
ΕΓΧΕΙΡΙΔΙΟ ΚΑΤΟΧΟΥ
Tensiomètre connecté (BPM1)
MANUEL D’UTILISATION
1
INTRODUCTION
PACKAGE CONTENTS
INTENDED USE
BLOOD PRESSURE CLASSIFICATION FOR ADULTS
CONTRAINDICATION
PARTS AND DISPLAY INDICATORS
OPERATING PROCEDURES AND SETUP
OPERATING PROCEDURES
Connecting The Cuff To The Monitor
Apply The Cuff
Body Posture
OPERATION INSTRUCTIONS
SETUP WIRELESS NETWORK
SPECIFICATIONS
GENERAL SAFETY AND PRECAUTIONS
BATTERY HANDLING AND USAGE
TECHNICAL ALARM DESCRIPTION
TROUBLESHOOTING
CARE AND MAINTENANCE
WARRANTY INFORMATION
EXPLANATION OF SYMBOLS
IMPORTANT INFORMATION REQUIRED BY THE FCC
OTHER STANDARDS AND COMPLIANCES
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table of Contents
EN
1
1
1
1
2
3
3
3
3
4
4
5
6
6
7
10
11
11
14
15
15
17
18
18
Wireless Blood Pressure Monitor (BPM1)
OWNER’S MANUAL
2
Note: This chart is not intended to provide a basis for any type of emergency condition or diagnosis
based on the color scheme; this chart only depicts different classifications of blood pressure. Consult
your physician for proper interpretation of blood pressure results.
CONTRAINDICATION
It is not recommended for people with serious arrhythmia to use this Blood Pressure Monitor.
Classification of blood pressure for adults
160
180
Systolic
(mmHg)
Diastolic
(mmHg)
140
130
120
80 85 90 100 110
Severe Hypertension
Moderate Hypertension
Mild Hypertension
High-normal BP
Normal BP
BLOOD PRESSURE
CLASSIFICATION SBP
mmHg DBP
mmHg
Optimal <120 <80
Normal 120-129 80-84
High-normal 130-139 85-89
Grade 1 Hypertension 140-159 90-99
Grade 2 Hypertension 160-179 100-109
Grade 3 Hypertension ≥=180 ≥=110
WHO/ISH Denitions and Classication of Blood Pressure Levels
3
Cuff tube plug
Charge plug
TFT display
User 1 button
User 2 button
OPERATING PROCEDURES AND SETUP
Connecting The Cuff To The Monitor
Insert the cuff tubing connector into the air port on the side of the
monitor. Make sure that the connector is completely inserted to avoid
air leakage during blood pressure measurements.
Avoid compression or restriction of the connection tubing during
measurement which may cause inflation error or harmful injury due to
continuous cuff pressure.
When the monitor is turned off, inserting the cuff tubing connector
will turn the monitor on.
PARTS AND DISPLAY INDICATORS
Cuff tubing connector Cuff
OPERATING PROCEDURES
INTRODUCTION
Thank you for selecting the iHealth Clear. The iHealth Clear is a fully automatic arm cuff blood pressure
monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The monitor
will transmit your measurements automatically to the iHealth cloud when it is connected to a wireless network.
PACKAGE CONTENTS
• 1 Blood Pressure Monitor
• 1 Quick Start Guide
• 1 Cuff
• 1 Charging Cable
INTENDED USE
The iHealth Clear (Electronic Sphygmomanometer) is intended for use in a professional setting or at
home and is a non-invasive blood pressure measurement system. It is designed to measure the systolic
and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an
inflatable cuff is wrapped around the upper arm. The measurement range of the cuff circumference is
8.6” to 16.5” (22 cm - 42 cm) or 16.5” to 18.9” (42 cm – 48 cm) with the optional cuff.
BLOOD PRESSURE CLASSIFICATION FOR ADULTS
The World Health Organization (WHO) has created the following guide for assessing high blood pressure
(without regard to age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.)
also need to be considered. Consult with your physician for accurate assessment.
4
1-2cm(1/2")
Apply The Cuff
a. Pull the cuff end through the metal loop, positioning it outward
(away from your body).
b. Place a bare arm through the cuff and position the cuff 1/2" (1-2 cm)
above the elbow joint.
c. Tighten the cuff and close it by pulling it towards your body, securing
it closed with the Velcro fastener.
d. While seated, place your hand palm-side up in front of you on a flat
surface such as a desk or table. Position the rubber tube in the
middle of your arm aligned with your middle finger.
e. The cuff should fit comfortably, yet snugly around your arm. You should be able to insert one finger
between your arm and the cuff.
Remember to:
1. Make sure that the appropriate cuff size is used; refer to the cuff circumference range in “SPECIFICATIONS”.
2. Measure on the same arm each time.
3. Stay still during measurement. Do not move your arm, body, or the monitor.
4. Stay still and calm for one to one and half minutes before taking a blood pressure measurement.
5. Keep the cuff clean. Cleaning the cuff after every 200 times of usage is recommended.
If the cuff becomes dirty, clean it with a moistened cloth. Do not allow any water to go into the interior
of the cuff, when the cuff is being cleaned. Do not rinse the monitor or cuff with running water.
Body Posture
Sitting Comfortably During Measurement
a. Be seated with your feet flat on the floor without crossing your legs.
5
Lying Down During Measurement
a. Lie on your back.
b. Place your arm straight along your side with your hand palm-side up.
c. The cuff should be placed at the same level as your heart.
Note: Blood pressure can be affected by the position of the cuff and your physiologic condition.
OPERATION INSTRUCTIONS
a. After applying the cuff and your body is in a comfortable position, press the User 1 or User 2 button to
turn the monitor on. The display will show "While measuring, please relax." The monitor will start to
find the baseline zero pressure. When the cuff is not connected with the monitor, press any key to
display indoor/outdoor temperature (if the wireless network is successfully set), battery level, and time
(if the wireless network is successfully set) for 3 seconds. The monitor will then turn off automatically.
b. The cuff will start to slowly inflate. The blood pressure and pulse will be measured during inflation.
Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed
on the screen. The result will automatically be stored in the Memory of the monitor. If memory is
b. Place your hand palm-side up in front of you on a flat surface such as a desk or table.
c. The cuff should be placed at the same level as your heart.
6
SPECIFICATIONS
1. Product name: iHealth Clear Wireless Blood Pressure Monitor
2. Reference: BPM1
3. Classification: Internally powered, Type BF applied part, IP20, No AP or APG, Continuous operation
4. Machine size: approx. 4.7” x 4.6” x 2.0” (119mm × 118mm × 51mm)
5. Cuff circumference: 8.6’’ - 16.5’’ (22 cm – 42 cm), 16.5’’-18.9’’ (42 cm – 48 cm) (Optional XL size)
6. Weight: approx. 12.3oz (350g) (excluding cuff)
7. Measuring method: Oscillometric method, automatic inflation and measurement
8. Memory volume: 2 × 1000 times with time and date stamp
9. Power: DC:5V 1.0A,
Battery: 1*3.7V Li-ion 2200mAh
stored in the current memory group, the monitor will show the difference between the current
measurement results and the previous measurement. If the monitor is connected to the cloud, the
measurement will be uploaded automatically.
c. After measurement, the monitor will turn off automatically after 1 minute of no operation.
d. During measurement, you can press the any key to turn off the monitor manually.
Important: Please consult a healthcare professional for interpretation of blood pressure measure-
ments
SETUP WIRELESS NETWORK
Download and install the iHealth MyVitals app from the Apple App Store or Google Play Store. Then
press and hold the User 2 button for 10 seconds. Follow the in app steps to finish setting up Wireless-
network.
GENERAL SAFETY AND PRECAUTIONS
1. Read all of the information in the Owner’s Manual and other provided instructions before operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
7
10. Measurement range:
Cuff pressure: 0-300 mmHg
Systolic: 60-260 mmHg
Diastolic: 40-199 mmHg
Pulse rate: 40-180 beats/minute
11. Accuracy:
Pressure: ±3 mmHg
Pulse rate: ±5%
12. Wireless communication:
IEEE 802.11b/g/n
13. Environmental temperature for operation: 10℃40℃ (50℉104℉)
14. Environmental humidity for operation: ≤85%RH
15. Environmental temperature for storage and transport: -20℃50℃ (-4℉122℉)
16. Environmental humidity for storage and transport: ≤85%RH
17. Environmental pressure: 80kPa-105kPa
18. Battery life: more than 180 measurements on a full charge
19. The blood pressure measurement system includes accessories, pump, valve, cuff, and sensor.
Note: These specifications are subject to change without notice.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same arm for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation in
your arm to recover. Prolonged over-inflation (cuff pressure exceed 300 mmHg or maintained above
15 mmHg for longer than 3 minutes) of the bladder may cause ecchymoma of your arm.
7. Consult your physician for any of the following situations:
a) The application of the cuff over a wound or inflamed area
b) The application of the cuff on any limb with intravascular access or therapy, or an arterio-venous (A-V) shunt
c) The application of the cuff on the arm on the side of a mastectomy
d) Simultaneous use with other medical monitoring equipment on the same limb
e) The blood circulation of the user needs to be checked
8. This Blood Pressure Monitor is designed for adults and should never be used on infants, young
children, pregnant or pre-eclamptic patients. Consult your physician before use on children.
9. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
10. Blood pressure measurements determined by this product are equivalent to those obtained by
professional healthcare practitioners using the cuff/stethoscope auscultation method within the limits
prescribed by the American National Standard, Electronic or Automated Sphygmomanometer.
11. Information regarding potential electromagnetic or other interference between the blood pressure
monitor and other devices together with advice regarding avoidance of such interference, please see
ELECTRO MAGNETIC COMPATIBILITY INFORMATION. It is suggested that the blood pressure
monitor be kept 10 meters away from other wireless devices, such as WLAN unit, cell phone,
microwave oven, etc.
12. Please do not use any other cuff other than that supplied by the manufacturer as this may result in
measurement errors and a biocompatible hazard.
13. This product might not meet its performance specifications if stored or used outside the specified
temperature and humidity ranges.
14. Please do not share the cuff with any infectious person to avoid cross-infection.
15. This product should not be used as a USB device.
16. If the determined blood pressure (systolic or diastolic) is outside the rated range specified in part
SPECIFICATIONS, the app will immediately display a technical alarm on screen. In this case, consult
a physician or ensure that proper measurement procedures are followed. The technical alarm is
preset in the factory and cannot be adjusted or inactivated. This technical alarm is assigned as low
priority according to IEC 60601-1-8.The technical alarm is non-latching and does not need to be
reset.
17. A medical AC adapter with an output of DC 5.0V and complies with IEC 60601-1/UL 60601-1 and
IEC 60601-1-2/EN 60601-1-2 is suitable for this monitor. Please note that the monitor jack size is
USB micro-B.
18. Use of Charging Cable other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
19. Measurements are not possible in patients with a high frequency of arrhythmias.
20. The device is not intended for use on neonates, children or pregnant women. (Clinical testing has
not been conducted on neonates, children or pregnant women.)
21. Motion, trembling, shivering may affect the measurement reading.
22. The device would not apply to the patients with poor peripheral circulation, noticeably low blood
pressure, or low body temperature (there will be low blood flow to the measurement position).
23. The device would not apply to the patients who use an artificial heart and lung (there will be no pulse).
9
BATTERY HANDLING AND USAGE
Do not replace the battery. If the battery can no longer be charged, please contact Customer Service.
• When charging is needed, please connect the monitor to a power source.
• When the cuff is not connected with the monitor, press any key to display indoor/outdoor temperature (if
wireless network is successfully set), battery level, and time (if wireless network is successfully set) for 3
seconds. If the battery level is less than 20% (the battery indicator will only display a battery symbol),
please charge the battery. The monitor will not work until the battery is sufficiently charged.
• When you charge the monitor, the monitor will display with different indicator indicating the charging
status. See the table below for details.
• It is suggested that you charge the battery when the battery is less than 25%.
Overcharging the battery may reduce its lifetime.
10
Lithium battery replacement by inadequately trained personnel could result in a hazard such as a fire
or explosion.
Do not plug or unplug the power cord into the electrical outlet with wet hands.
If the AC adapter is abnormal, please change the adapter.
Do not use the monitor while charging.
Do not use any other type of AC adapter as it may harm the monitor.
24. Consult your physician before using the device for any of the following conditions: common
arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor
perfusion, diabetes, pre-eclampsia, enal diseases.
25. The patient can be an intended operator.
26. Please contact the service center if the character or number displayed is incomplete.
11
The monitor, cable, battery and cuff must be disposed of according to local regulations at the end
their usage.
NoteThe battery has limited charge cycles and may eventually need to replaced by an iHealth service
provider. Battery life and charge cycles vary by use and settings.
Monitor Status
Charging
Fully charged
Status Indicator
Dynamical battery symbol
Charging battery symbol
Empty battery symbol and reminder: "Low battery, unable to perform
measurement, please recharge""
Low battery
TECHNICAL ALARM DESCRIPTION
The monitor will show SYS or DIA beyond measurement range, or SYS or DIA below measurement range as
technical alarm on screen with no delay if the determined blood pressure (systolic or diastolic) is outside the
rated range specified in part SPECIFICATIONS. In this case, you should consult a physician or check if your
operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or
deactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and needs no reset. The signal displayed on screen will disappear
automatically after about 8 seconds.
TROUBLESHOOTING
PROBLEM POSSIBLE CAUSE SOLUTION
Low Battery Battery is less than 20% Charge the battery
12
The cuff position was not correct or
it was not properly tightened
Body posture was not correct
during testing
Speaking, moving arm or body,
being angry, excited or nervous
during test
Incorrect operation or strong
electromagnetic interference
Review the cuff supplication instructions
and retest
Review body posture instructions and
retest
Retest when calm; avoid speaking
or movement during the test
Press the User 2 button for 10 seconds
to reset the device and try again.
No response
Display reads an
abnormal result
CARE AND MAINTENANCE
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result
in damage to the monitor.
3. If this monitor is stored near freezing temperatures, allow it to acclimate to room temperature before
use.
4. Do not attempt to disassemble this monitor.
5. If the monitor is not used for a long time, please be sure to fully charge it every month.
6. It is recommended that product performance be checked every 2 years or after each repair. Please
contact the service center.
7. No monitor component needs to be maintained by the user. The circuit diagrams, component part
lists,descriptions, calibration instructions, or other information which will assist the user’s appropri-
ately qualified technical personnel to repair those parts of the equipment which are designated for
repair can be supplied.
8. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water, diluted
disinfectant alcohol, or diluted detergent
9. The monitor can maintain the safety and performance characteristics for a minimum of 10,000
measurements or three years of usage, and the cuff integrity is maintained after 1,000 open–close
cycles of the closure.
10. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.
Battery replacement should only be performed by a qualified iHealth technician. To do otherwise will
void your warranty and possibly damage your unit.
11. It is recommended that if the cuff is used, for example, in a hospital or a clinic, it be disinfected
1413 15 16
WARRANTY INFORMATION
The Blood Pressure Monitor is warranted to be free from defects in materials and workmanship within
one year from the date of purchase when used in accordance with the provided instructions. The
warranty extends only to the end user. We will, at our option, repair or replace without charge the Blood
Pressure Monitor covered by the warranty. Repair or replacement is our only responsibility and your only
remedy under the warranty.
EXPLANATION OF SYMBOLS
Symbol for ”THE OPERATION GUIDE MUST BE READ”
The sign background color: blue. The sign graphical symbol: white.
Symbol for “WARNING”
Symbol for “TYPE BF APPLIED PARTS” (The cuff is type BF applied part)
Symbol for “MANUFACTURER”
Symbol for “COMPILES WITH MDD93/42/EEC REQUIREMENTS”
Symbol for“ DATE OF MANUFACTURE”
Symbol for “SERIAL NUMBER”
Symbol for “EUROPEAN REPRESENTATIVE”
Symbol for “KEEP DRY”
iHealth is a trademark of iHealth Labs, Inc.
Manufactured for iHealth Labs, Inc.
120 San Lucar Ct.,Sunnyvale, CA 94086, USA
1-855-816-7705
www.ihealthlabs.com
iHealthLabs Europe SARL
3 rue Tronchet, 75008, Paris, France
ANDON HEALTH CO., LTD.
No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190, China
Tel: 86-22-60526161
twice a week. Wipe the inner side (the side that contacts skin) of the cuff with a soft cloth lightly
moistened with Ethyl alcohol (75-90%). Then air dry the cuff.
support@ihealthlabs.eu www.ihealthlabs.eu
Symbol for “DO NOT TOUCH THE CONNECTOR WITH ESD
WARNING SYMBOL DURING MEASUREMENT”
17
IMPORTANT INFORMATION REQUIRED BY THE FCC
This device complies with Part 15 of the FCC Rules. Its operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by iHealth Labs, Inc. would void the user’s authority
to operate the product.
NOTE: This product has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this product does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
— Increase the separation between the equipment and receiver.
— Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
— Consult the dealer or an experienced radio/TV technician for help.
The SAR limit of USA (FCC) is 1.6 W/kg averaged over one gram of tissue. Device has also been tested
against this SAR limit.
This product complies with Industry Canada. IC: RSS-210
IC NOTICE
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the
following two conditions:
OTHER STANDARDS AND COMPLIANCES
The Blood Pressure Monitor corresponds to the following standards: IEC 60601-1:2005
+A1:2012(E)/EN 60601-12006/A11: 2011 (Medical electrical equipment -- Part 1: General require-
ments for basic safety and essential performance), IEC 60601-1-2:2014(Medical electrical equipment --
Part 1-2: General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests), IEC80601-2-30:2009+AMD12013/EN
80601-2-30:2010/A12015(Medical electrical equipment –Part 2-30: Particular requirements for the
basic safety and essential performance of automated non-invasive sphygmomanometers)EN 1060-1:
1995 + A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997
+ A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-me-
chanical blood pressure measuring systems).
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This product is applicable to the equipment and system requirements for the purpose of receiving radio
frequency energy for the purpose of the work,Wireless network receive bandwidth 22M. This product
can also be used to include RF transmitter equipment and system requirements and emission frequency
of 2.4GHz ISM band, Wireless network modulation types: DBPSK/DQPSK/CCK (DSSS)
BPSK/QPSK/16QAM/64QAM (OFDM) , effective radiated power: < 20dBm .
18
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation
of the device.
This product is approved in accordance to R&TTE directive transmitter.
Hereby, [iHealth Labs, Inc.],declares that this BPM1 is in compliance with the essential requirements and
other relevant provisions of Directive 1999/5/EC. Directive 1999/5/EC declaration of conformity can be
downloaded on the following link: https://www.ihealthlabs.eu/support/certifications
Table 1 - Emission
Phenomenon
RF emissions
Harmonic distortion
Voltage fluctuations and flicker
CISPR 11 Group 1, Class B
IEC 61000-3-2 Class A
IEC 61000-3-3 Compliance
Electromagnetic environment
Home healthcare environment
Home healthcare environment
Home healthcare environment
Compliance
Table 2 - Enclosure Port
Phenomenon
Electrostatic Discharge
Radiated RF EM field
Proximity fields from RF
wireless communications
equipment
Rated power frequency
magnetic fields
Basic EMC standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air
10V/m 80MHz-2.7GHz 80% AM at 1kHz
Refer to table 3
30A/m 50Hz or 60Hz
Immunity test levels
Home Healthcare Environment
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency (MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Band (MHz)
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
Immunity test levels
Professional healthcare facility environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
19 20
Table 4 – Input a.c. power Port
Phenomenon
Electrical fast
transients/burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted
disturbances
induced by RF fields
Voltage dips
Voltage interruptions
Basic EMC standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Immunity test levels
Home Healthcare Environment
±2 kV 100kHz repetition frequency
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV, ±2 kV
3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands between
0.15MHz and 80MHz
80%AM at 1kHz
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0º
0% UT; 250/300 cycles
21
Symbol for “ENVIRONMENT PROTECTION – Waste electrical products should not be disposed
of with household waste. Please recycle where facilities exist. Check with your local authority or
retailer for recycling advice”.
LCD shows
“Error Code 000”
LCD shows
“Error Code 001”
LCD shows
“Error Code 002”
LCD shows
“Error Code 003”
LCD shows
“Error Code 004”
LCD shows
“Error Code 005”
LCD shows
“Error Code 006”
LCD shows
“Error Code 007”
LCD shows
“Error Code 010”
LCD shows
“Error Code 011”
Pressure system is unstable before
measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff is
too tight during inflation
Pneumatic system leakage or cuff is
too loose during inflation
Cuff pressure above 300mmHg
More than 160 seconds with cuff
pressure above 15 mmHg
Memory abnormality or temperature
chip error
Pressure sensor parameter error
When uploading data not connected
to the cloud
Don’t move and try again.
Apply the cuff correctly and try again
Measure again after five minutes.
If the monitor is still abnormal,
please contact the local distributor
or the factory.
LCD shows
“Error Code 012”
LCD shows
“Error Code 013”
LCD shows
“Error Code 015”
LCD shows
Wireless network
password is wrong”
Wireless network
module initialization failed
Network not found
Failed to get IP
Wireless network password is wrong
Re-set Wireless network and make
sure the connected network is in good
condition. Measure again after five
minutes. If the monitor is still abnormal,
please contact the local distributor or
the factory.
Check the Wireless network password
and re-set Wireless network


Produkt Specifikationer

Mærke: IHealth
Kategori:
Model: BPM1 - Clear

Har du brug for hjælp?

Hvis du har brug for hjælp til IHealth BPM1 - Clear stil et spørgsmål nedenfor, og andre brugere vil svare dig




IHealth Manualer