Kyoritsu KEW 8146 Manual
Kyoritsu
Måleudstyr
KEW 8146
Læs nedenfor 📖 manual på dansk for Kyoritsu KEW 8146 (1 sider) i kategorien Måleudstyr. Denne guide var nyttig for 9 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere
Side 1/1
INSTRUCTIONMANUAL
KEW 8146/8147/8148
LEAKAGE CLAMP SENSOR
Series
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
KEW
8146
KEW
8147
KEW
8148
LEAKAGECLAMPSENSOR
1. Safety warnings
This instrument has been designed, manufactured and
tested according to IEC
61010
: Safety requirements for
Electronic Measuringapparatus, and deliveredin the best
condition after passing quality control tests. This
instruction manual contains warnings and safety rules
which have to be observed by the user to ensure safe
operation of the instrument and to maintain it in safe
condition. Therefore, read through these operating
instructionsbeforeusingtheinstrument.
# WARNING
●
Read through and understand instructionscontained in
thismanualbeforeusingtheinstrument.
●
Keep the manual at hand to enable quick reference
whenevernecessary.
●
The instrument is to be used only in its intended
applications.
Theoperatinginstructionsdescribedinthemanualmust
beobserved.
●
Understand and follow all the safety instructions
containedinthemanual.
Itisessentialthattheaboveinstructionsareadheredto.
Failuretofollowtheaboveinstructionsmaycauseinjury
andorinstrumentdamage.
Kyoritsuisbynomeansliableforanydamageresulting
from the instrument in contradiction to this cautionary
note.
The symbol
#
indicated on the instrument, means that
theusermustrefertotherelatedpartsinthemanualfor
safeoperationoftheinstrument.Itisessentialtoreadthe
instructions wherever the
#
symbol appears in the
manual.
#
DANGER
isreservedforconditionsandactionsthat
arelikelytocauseseriousorfatalinjury.
#
WARNING
is reserved for conditions and actions
thatcancauseseriousorfatalInjury.
#
CAUTION
isreservedforconditionsandactionsthat
cancauseminorinjuryorinstrumentdamage.
# DANGER
●
Never make measurement on a circuit in which the
electricalpotentialexceedsAC
300
V.
●
Donot makemeasurementwhenthunderrumbling. Ifthe
instrument is in use, stop the measurement immediately
andremovetheinstrumentfromthemeasuredobject.
●
Donotattempttomakemeasurementinthepresenceof
flammable gasses. Otherwise, the use ofthe instrument
maycausesparking,whichcanleadtoanexplosion.
●
The transformer jaws are made of metal and their tips
are not completely insulated. Be especially careful
aboutthepossible shortingwhere themeasuredobject
hasexposedmetalparts.
●
Never attempt to use the instrument if it's surface or
yourhandarewet.
●
Do not exceed the maximum allowable input of any
measuringrange.
# WARNING
●
Never attempt to make any measurement, if any
abnormal conditions are noted, such as broken case,
andexposedmetalparts.
●
Donotinstallsubstitutepartsormakeanymodification
totheinstrument.
Return the instrument to the distributor from who you
purchased this instrument for repair or re-calibration in
caseofsuspectedfaultyoperation.
●
Always keep your fingers and hands behind the barrier
ontheinstrumenttoavoidthepossibleshockhazard.
# CAUTION
●
Donotsteponorpinchthecordtopreventthejacketof
cordfrombeingdamaged.
●
The output connector shall be removed or connected
without clamping a conductor. Otherwise, it may cause
afailure.
●
Do not expose the instrument to direct sunlight, high
temperatures,humidityordew.
●
Nevergiveshocks,suchasvibrationordrop,whichmay
damagetheinstrument.
●
Use a damp cloth and detergent for cleaning the
instrument.Donotuseabrasivesorsolvents.
Safetysymbols
Incoming wire
O:Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring
○
MeasurementCategory:
To ensure safe operation of measuring instruments,
IEC
61010
establishes safety standards for various
electrical environments, categorized as O to CAT IV,
and called measurement categories. Higher-numbered
categories correspond to electrical environments with
greater momentary energy, so a measuring instrument
designed for CAT III environments can endure greater
momentaryenergythanonedesignedforCATII.
O :Circuitswhicharenotdirectlyconnectedtothe
mainspowersupply.
CATII :Electricalcircuitsofequipmentconnectedtoan
ACelectricaloutletbyapowercord.
CATIII :Primary electrical circuits of the equipment
connecteddirectlyto thedistribution panel, and
feedersfromthedistributionpaneltooutlets.
CATIV :The circuitfrom the service drop to the service
entrance, and to the power meter and primary
over-current protection device (distribution
panel).
2. Features
●
ClampsensorforACleakagecurrentmeasurement.
●
Canmeasureupto:
KEW
8146
:
30
A
KEW
8147
:
70
A
KEW
8148
:
100
A
●
DesignedtointernationalsafetystandardIEC
61010
-
2
-
032
CAT
Ⅲ
PollutionDegree
2
3. Instrument layout
TransformJaws Trigger Cable
Outputconnector
Barrier
Barrier
:
It is a part providing protectionagainst electrical
shock and ensuring the minimum required air and
creepagedistances.
4. DIN Plug pin assignment
3
:GNDpin
5
:Outputsignalpin
6
:SensorSignalpin
(
Resistancebetween
3
Pinand
6
Pin:
8146
:
47
k
Ω
8147
:
8
.
2
k
Ω
8148
:
30
k
Ω)
1
,
2
and
4
:Nouse
*Lower figure shows the pin assignment seeing the Clamp
sensor from output connector part. The figure of the pin
assignment of connection terminal is symmetrical to lower
figure.
5
3
12
4
6
5. Operating instructions
# DANGER
●
Never make measurement on a circuit in which the
electricalpotentialexceedsAC
300
Vinordertoavoid
possibleshockhazard.
●
Thetransformerjawsaremadeofmetalandtheirtips
are not completely insulated. Be especially careful
aboutthepossibleshortingwherethemeasuredobject
hasexposedmetalparts.
●
Keepyourfingersandhandsbehind thebarrierduringa
measurement.
# CAUTION
●
Take sufficient care to avoid shock, vibration or
excessive force when handling the instrument.
Otherwise, precisely adjusted transformer jaws willbe
damaged.
●
Whentransformerjawsdonotfullyclose,nevertryto
closethembyforce,butmakethemfreetomoveand
try again. If a foreign substance is stuck in the jaw
tips,removeit.
●
When making current measurements, keep the
transformerjawsfullyclosed.
Otherwise,accuratemeasurementscannotbetaken.
●
Hold the inserting part (except for the cable) and
disconnect the Output connector from the measuring
instrumentsoasnottocauseabreakinthecord.
●
When measuring current which pulse element is
superposed,differencesoftheindicatedvaluemaybe
causedbetweenranges,ifthepeakvalueexceedsthe
measurementrangetoalargeextent.Inthis
case,the
reading at the bigger range should be taken
as a right
value.SensitivetransformerjawsareusedforLeakage
clamp meter. Because of the characteristics of
transformerjaws,whichcanbeopenedandclosed,it
isimpossible to eliminate the interference of external
magnetic field completely. If there is a presence of
strongmagneticfield,usetheinstrumentatadistance
as far as possible from it. Following are the typical
thingsgeneratingmagneticfield.
・
Conductorfedlargecurrent
・
Motor
・
Equipmentwhichhasmagnet
・
Integratingwattmeter
5
-
1
Measuringmethod
(
1
)
ConnecttheOutputconnectortotheInputterminalof
themeasuringinstrument.
(
2
)Press the Trigger to open the transformer jaws and
clampontooneconductor.
Inthiscase,themeasuredconductorshallbeatthe
centerofthejaws.
(
3
)
Ensurethatthetipsoftransformerjawsarefirmlyclosed.
5
-
2
Measuringmethod
(
1
)Measuring out of balance leakage current (See
Fig.
1
):Clamp onto all conductors except a grounded
wire.
(
2
)Measuring earth leakage current (See Fig.
2
):Clamp
ontoagroundedwire.
3
-phase
3
-wiresystem
(In
4
-wiresystemwithneutral,
clampontoall
4
wires)
Fig.
1
Measuringoutofbalanceleakagecurrent
Groundedwire
Load
Load
Load
Single-phase
2
-wiresystem
(In
3
-wiresystemwithneutral,
clampontoall
3
wires)
Fig.
2
Measuringearthleakagecurrent
92
-
1746
E
7
-
22
DISTRIBUTOR
#
Refertotheinstructionsinthemanual.
IndicatesaInstrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on live bare
conductors when the voltage to be tested is below
Circuit -Ground-to-Earth voltage againsttheindicated
MeasurementCategory.
~
IndicatesAC
Model KEW
8146
KEW
8147
KEW
8148
Ratedvoltage AC
30
Arms(
42
.
4
Apeak) AC
70
Arms(
99
.
0
Apeak) AC
100
Arms(
141
.
4
Apeak)
Outputvoltage AC
0
〜
1500
mV
(
50
mV/A)
AC
0
〜
3500
mV
(
50
mV/A)
AC
0
〜
5000
mV
(
50
mV/A)
Measuringrange AC
0
〜
30
A AC
0
〜
70
A AC
0
〜
100
A
Accuracy
(Input:sinewave)
0
A
〜
15
A
±
1
.
0
%rdg
±
0
.
1
mV(
50
/
60
Hz)
±
2
.
0
%rdg
±
0
.
2
mV(
40
〜
1
kHz)
15
A
〜
30
A
±
5
.
0
%rdg(
50
/
60
Hz)
±
10
.
0
%rdg(
45
〜
1
kHz)
0
A
〜
40
A
±
1
.
0
%rdg
±
0
.
1
mV(
50
/
60
Hz)
±
2
.
0
%rdg
±
0
.
2
mV(
40
〜
1
kHz)
40
A
〜
70
A
±
5
.
0
%rdg(
50
/
60
Hz)
±
10
.
0
%rdg(
45
〜
1
kHz)
0
A
〜
80
A
±
1
.
0
%rdg
±
0
.
1
mV(
50
/
60
Hz)
±
2
.
0
%rdg
±
0
.
2
mV(
40
〜
1
kHz)
80
A
〜
100
A
±
5
.
0
%rdg(
50
/
60
Hz)
±
10
.
0
%rdg(
45
〜
1
kHz)
Temperature&humidityrange
(Guaranteedaccuracy)
23
±
5
℃
,relativehumidity:
85
%orless(nocondensation)
Operatingtemperaturerange
0
〜
50
℃
,relativehumidity:
85
%orless(nocondensation)
Storagetemperaturerange
-
20
〜
60
℃
,relativehumidity:
85
%orless(nocondensation)
Maximumpermissibleinput
AC
30
Armscontinuous(
50
/
60
Hz) AC
70
Armscontinuous(
50
/
60
Hz) AC
100
Armscontinuous(
50
/
60
Hz)
Output
impedance Approx.
90
Ω
Approx.
100
Ω
Approx.
60
Ω
Locationforuse Altitudeupto
2000
m,Indoors
Applicable
standards
IEC
61010
-
1
,IEC
61010
-
2
-
032
MeasurementCAT
Ⅲ
(
300
Vrms)
Pollutiondegree
2
IEC
61326
-
1
(EMC)
Environmental
standards EURoHSdirectivecompliant
Withstand
voltage
AC
3470
Vrms(
50
/
60
Hz)for
5
sec.
betweenJawandenclosure
betweenenclosureandoutputterminal
betweenJawandoutputterminal
Insulation
resistance
50
M
Ω
orgreaterat
1000
V
betweenJawandenclosure
betweenenclosureandoutputterminal
betweenJawandoutputterminal
ConductorSize Approx.
24
mmindiameter(max.) Approx.
40
mmindiameter(max.) Approx.
68
mmindiameter(max.)
Dimension
100
(L)
×
60
(W)
×
26
(D)mm
128
(L)
×
81
(W)
×
36
(D)mm
186
(L)
×
129
(W)
×
53
(D)mm
Cablelength Approx.
2
m
Outputterminal MINIDIN
6
PIN
Weight Approx.
150
g Approx.
240
g Approx.
510
g
Accessories Instructionmanual
Cablemarker
Option MODEL
7146
(
Banana
Φ
4
adjusterplug
)
MODEL
7185
(
Extensioncable
)
6. Specifications
Produkt Specifikationer
Mærke: | Kyoritsu |
Kategori: | Måleudstyr |
Model: | KEW 8146 |
Har du brug for hjælp?
Hvis du har brug for hjælp til Kyoritsu KEW 8146 stil et spørgsmål nedenfor, og andre brugere vil svare dig
Måleudstyr Kyoritsu Manualer
2 September 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
Måleudstyr Manualer
- Måleudstyr Bosch
- Måleudstyr SilverCrest
- Måleudstyr Beurer
- Måleudstyr Panasonic
- Måleudstyr Hager
- Måleudstyr Owon
- Måleudstyr DeWalt
- Måleudstyr Extech
- Måleudstyr Silverline
- Måleudstyr IFM
- Måleudstyr Victron Energy
- Måleudstyr Pioneer
- Måleudstyr Finder
- Måleudstyr Mitsubishi
- Måleudstyr Metrix
- Måleudstyr ProfiCare
- Måleudstyr TFA
- Måleudstyr Soler & Palau
- Måleudstyr Nedis
- Måleudstyr Milwaukee
- Måleudstyr BENNING
- Måleudstyr Wago
- Måleudstyr Fysic
- Måleudstyr Microlife
- Måleudstyr Topcom
- Måleudstyr Laserliner
- Måleudstyr Trotec
- Måleudstyr Einhell
- Måleudstyr Honeywell
- Måleudstyr Megasat
- Måleudstyr Soehnle
- Måleudstyr REV
- Måleudstyr Carrier
- Måleudstyr Danfoss
- Måleudstyr Eberle
- Måleudstyr Bresser
- Måleudstyr Goobay
- Måleudstyr Digitus
- Måleudstyr Alecto
- Måleudstyr Ansmann
- Måleudstyr Westfalia
- Måleudstyr Homedics
- Måleudstyr Basetech
- Måleudstyr Lanaform
- Måleudstyr Grohe
- Måleudstyr Voltcraft
- Måleudstyr Testboy
- Måleudstyr Omnitronic
- Måleudstyr Velleman
- Måleudstyr FLIR
- Måleudstyr Perel
- Måleudstyr Vemer
- Måleudstyr ORNO
- Måleudstyr Amprobe
- Måleudstyr Technoline
- Måleudstyr Fluke
- Måleudstyr Stabila
- Måleudstyr Shimano
- Måleudstyr Laica
- Måleudstyr Techno Line
- Måleudstyr Ridgid
- Måleudstyr Midland
- Måleudstyr Monacor
- Måleudstyr Elgato
- Måleudstyr Hazet
- Måleudstyr LogiLink
- Måleudstyr Testo
- Måleudstyr Be Cool
- Måleudstyr GW Instek
- Måleudstyr GoClever
- Måleudstyr LaCie
- Måleudstyr Rossmax
- Måleudstyr Ideal
- Måleudstyr Multimetrix
- Måleudstyr Eurolite
- Måleudstyr Eltako
- Måleudstyr Tempest
- Måleudstyr Meec Tools
- Måleudstyr Brüder Mannesmann
- Måleudstyr Eastron
- Måleudstyr Clean Air Optima
- Måleudstyr Vimar
- Måleudstyr Continental Edison
- Måleudstyr Klein Tools
- Måleudstyr Oregon Scientific
- Måleudstyr PeakTech
- Måleudstyr Dahua Technology
- Måleudstyr Uni-T
- Måleudstyr Reely
- Måleudstyr Joy-It
- Måleudstyr HT Instruments
- Måleudstyr Kübler
- Måleudstyr TrueLife
- Måleudstyr Kern
- Måleudstyr Chauvin Arnoux
- Måleudstyr Kogan
- Måleudstyr Delta Ohm
- Måleudstyr Megger
- Måleudstyr Peak
- Måleudstyr Werma
- Måleudstyr Kemo
- Måleudstyr TDE Instruments
- Måleudstyr H-Tronic
- Måleudstyr Holzmann
- Måleudstyr ClimeMET
- Måleudstyr Adwa
- Måleudstyr Gossen Metrawatt
- Måleudstyr Lund
- Måleudstyr Yato
- Måleudstyr Beha-Amprobe
- Måleudstyr Cocraft
- Måleudstyr Emko
- Måleudstyr Hikmicro
- Måleudstyr Sonel
- Måleudstyr Murideo
- Måleudstyr Seaward
- Måleudstyr SRS
- Måleudstyr Toolland
- Måleudstyr VirtuFit
- Måleudstyr AS - Schwabe
- Måleudstyr IHealth
- Måleudstyr Fantini Cosmi
- Måleudstyr Sekonic
- Måleudstyr Eizo
- Måleudstyr Wentronic
- Måleudstyr Comark
- Måleudstyr Livington
- Måleudstyr Maximum
- Måleudstyr Tektronix
- Måleudstyr SIIG
- Måleudstyr Muller
- Måleudstyr TFA Dostmann
- Måleudstyr Greisinger
- Måleudstyr YSI
- Måleudstyr Kreg
- Måleudstyr Pancontrol
- Måleudstyr X-Rite
- Måleudstyr Ebro
- Måleudstyr ENTES
- Måleudstyr Arya
- Måleudstyr RK Technology
- Måleudstyr Aktobis
- Måleudstyr KS Tools
- Måleudstyr Trumeter
- Måleudstyr Aeros
- Måleudstyr Aqua Master Tools
- Måleudstyr Hanna
- Måleudstyr Atmel
- Måleudstyr Qualita
- Måleudstyr MGL Avionics
- Måleudstyr AREXX
- Måleudstyr PCE Instruments
- Måleudstyr UHoo
- Måleudstyr CEM
- Måleudstyr SHX
- Måleudstyr Hameg
- Måleudstyr CESYS
- Måleudstyr Circutor
- Måleudstyr Wachendorff
- Måleudstyr Duro Pro
- Måleudstyr Aerospace Logic
- Måleudstyr Zircon
- Måleudstyr BJZ
- Måleudstyr Johnson Level
- Måleudstyr GMW
- Måleudstyr SOEKS
- Måleudstyr Apollo Ultrasonic
- Måleudstyr Softing
- Måleudstyr Weltron
- Måleudstyr Oscium
- Måleudstyr Janitza
- Måleudstyr Sanwa
- Måleudstyr Condtrol
- Måleudstyr J.P. Instruments
- Måleudstyr Weidmüller
- Måleudstyr Qubino
- Måleudstyr Lascar Electronics
- Måleudstyr P3 International
- Måleudstyr Beckmann & Egle
- Måleudstyr Seek Thermal
- Måleudstyr INFICON
- Måleudstyr Motrona
- Måleudstyr Tempo
- Måleudstyr Camille Bauer
- Måleudstyr Blebox
- Måleudstyr HOZO Design
- Måleudstyr Corentium
- Måleudstyr GQ
- Måleudstyr Dostmann Electronic
- Måleudstyr SecuTech
- Måleudstyr NetPeppers
- Måleudstyr SensoProtect
- Måleudstyr Water-i.d.
- Måleudstyr ZKETECH
- Måleudstyr LabNation
- Måleudstyr Electronics International
- Måleudstyr Cliff
- Måleudstyr Fixpoint
- Måleudstyr AkYtec
- Måleudstyr Stahlwille
- Måleudstyr MASIMO
- Måleudstyr Milesight
- Måleudstyr Enda
- Måleudstyr Ermenrich
- Måleudstyr Noyafa
- Måleudstyr Stamos
Nyeste Måleudstyr Manualer
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024