Installation Manual
Type:
:
:
:: Medical Switching Power Supply
(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)
Installation
(1) Enclosed Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Enclosed units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Enclosed units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
(2) Modular Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Modular units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Modular units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
Produkt Specifikationer
Mærke: | Mean Well |
Kategori: | ikke klassificeret |
Model: | RPD-60A |
Har du brug for hjælp?
Hvis du har brug for hjælp til Mean Well RPD-60A stil et spørgsmål nedenfor, og andre brugere vil svare dig
ikke klassificeret Mean Well Manualer
16 August 2024
12 August 2024
8 August 2024
7 August 2024
ikke klassificeret Manualer
- ikke klassificeret Hager
- ikke klassificeret LG
- ikke klassificeret LiftMaster
- ikke klassificeret Sanus
- ikke klassificeret Auto Joe
- ikke klassificeret Shimano
- ikke klassificeret TC Electronic
- ikke klassificeret Dormakaba
- ikke klassificeret Martin
- ikke klassificeret Eurolite
- ikke klassificeret Polaris
- ikke klassificeret ATen
- ikke klassificeret Microchip
- ikke klassificeret Lindy
- ikke klassificeret Icy Box
- ikke klassificeret WMD
- ikke klassificeret Novation
- ikke klassificeret Urbanista
- ikke klassificeret Doepfer
- ikke klassificeret Lifan
Nyeste ikke klassificeret Manualer
21 August 2024
21 August 2024
21 August 2024
21 August 2024
19 August 2024
19 August 2024
19 August 2024
19 August 2024
18 August 2024
16 August 2024