Rossmax NK1000 Manual


Læs nedenfor 📖 manual på dansk for Rossmax NK1000 (16 sider) i kategorien Ikke kategoriseret. Denne guide var nyttig for 9 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/16
FCleaning procedures
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules may
result in damage, inecient nebulization or aspiration or infection.
Before use or after each treatment, the manufacturer advises you to clean and/or disinfect the
unit. Make sure the power switch of the nebulizer has been turned to the “OFF (O) position and
the unit has been disconnected from the power source before any cleaning procedures.
Disconnect the air tube from both nebulizer kit/nasal washer/nasal aspirator and nebulizer device
after each use and disassemble the parts of the unit described as below.
F-1 Nebulizer kit
1. Gently twist and pull up the cover of the nebulizer kit to open and separate.
2. Always dispose of any remaining medication in the medication cup after each use.
F-2 Nasal washer
1. Detach nasal adaptor, top cover, separator and nozzle from the unit.
2. Dispose of the solutions in the return uid collection chamber after each use.
F-3 Nasal aspirator
1. Detach the silicone tip with tube, top cover and lter from the unit.
2. Dispose of the mucus in the collection chamber after each use.
Sanitization
Sanitize and rinse each part with clean, warm water (about 40 C) with gentle dishwashing liquid °
(non-abrasive). After cleaning, leave the parts to dry in an open, clean environment.
Disinfection
The accessories which can be disinfected are nebulizer kit, mouthpiece, nasal washer and nasal
aspirator. You can choose one of the methods described below.
Method A: wash with a commercial (bacterial-germicidal) disinfectant and follow the instructions
and dilution rates supplied by the disinfectant manufacturer carefully.
Method B: boil each of the parts in water for 15~20 minutes, it is preferable to use de-mineralized
or distilled water to prevent calcium deposits.
It is eective on the parts that undergo this treatment only if the parts to be treated have previously
been sanitized. After cleaning, leave the parts to dry in an open, clean environment.
Note:
For the rst time cleaning or after the unit has been stored for an extended period of time,
thoroughly clean all components, except the air tube. The nebulizer kit is dishwasher safe.
• Adult and Child mask can only be washed at <40°C(104°F)
• Never clean with benzene, thinner or a ammable chemical.
After washing with the commercial disinfectant, please rinse with clean warm water and dry
thoroughly before using.
Cleaning the compressor
Wipe the compressor daily using a soft cloth.
NOTE: Any other form of cleaning or cleaning agents may damage the nish of the unit.
GChanging the air lter <Nebulizer device>
It is important to change the air lter approximately when the air lter turns gray. It is recommended
to change air lter every 2 months.
1. Remove the air lter cover by gently pulling forward.
2. Discard the gray lter.
3. Replace with a new, clean air lter.
4. Securely re-attach the air lter cover to the unit.
Troubleshooting
If any abnormality occurs during use, please check and correct the followings. We recommend
that you consult your local distributor or authorized Rossmax service center if, after checking the
described conditions below the device is still not working properly.
Problem Possible cause/Remedy
Unit does not operate when
power switch is pressed. Check the AC connection to the power outlet.
No nebulization/aspiration or lack
of nebulization/aspiration, when
the power is on.
Check the main unit if there is any physical damage.
Check the operating pressure/vacuum meets the
requirement.
Check the air lter and replace if necessary.
Check the tubing is not clogged, twisted or broken.
Make sure that the air tube is correctly connected to the
compressor and nebulizer kit or nasal washer or nasal
aspirator.
Check the parts are clean, If not, carry out the cleaning
procedures.
Nebulizer kit/ nasal washer/ nasal
aspirator does not work
Check the nozzle is set properly in place inside the nebulizer
kit/ nasal washer.
Reassemble each individual part and make sure it is in good
condition.
Check that there is medication in the nebulizer cup.
Make sure that the nebulizer kit is tilted at an angle less than
45°, or the aspirator is upright.
Check the solution has been properly poured into the uid
chamber inside the nasal washer.
Check the trigger button has been pressed completely.
The compressor is very hot
Do not cover the compressor during use.
Do not use the device in an environment with temperatures
higher than 40°C.
Protection against electric shock:
• Class II equipment.
Type BF applied parts:
• Mouthpiece, mask, nasal washer and nasal aspirator.
Protection against harmful ingress of water and particulate matter:
• IP21
Degree of safety in the presence of ammable anesthetics or oxygen:
No AP/APG (not suitable for use in the presence of ammable anesthetics or oxygen).
E-3 Operating your nasal aspirator
Dry and very thick mucus inside the nose may be dissolved by saline solution. Nasal washer is
recommended to be used before using nasal aspirator to help remove nasal mucus and secretions.
1. Turn the power switch to the “Nasal Aspirator” (=) position.
2. Place your index nger over the opening on the nasal aspirator to activate the aspiration.
3. Adjust the level of vacuum turning vacuum regulator, and check that the unit makes the vacuum
without any leakage and this level varies when the vacuum regulator is acted on.
Note: Turn the vacuum regulator clockwise, in the direction, to obtain a higher vacuum; or “+”
anti-clockwise for a lower vacuum.
4. Insert the silicone tip into one of the nostrils gently. Use the aspirator for 3 seconds for 4 times
per nostril.
Caution:
• Before using the nasal aspirator, try it on the hand to see if the aspiration works.
The nasal aspirator is applicable to baby. Keep your baby in a comfortable position and make
sure that his/her head is lifted.
While using, keep the aspirator in the uprigh t position with the nger opening on top, and the
collection chamber should be emptied when the stem inside the unit is overlled with sucked
mucus. This prevents the mucus from entering the suction pump.
• Do not tilt or place the nasal aspirator on the table before turning o the device.
Note:
• Air lters cannot be cleaned or washed.
Only NK1000 air lters can be used. Do not substitute alternate material such as cotton.
• Do not operate without an air lter.
Do not expose air lter to direct sunlight, or the performance is impaired due to aging phenomenon.
• Check that the air lters are clean and free of dust before inserting them.
Changing the lter <Nasal aspirator>
The lter supplied with Rossmax nasal aspirator is recommended to be replaced for each new user,
or if overow occurs or the lter is saturated. It should also be replaced if the unit is used with
dierent person whose risk of contamination is unknown.
Note:
The lter cannot be washed, sanitized, disinfected, or sterilized.
Do not use the nasal aspirator without the lter, the suction pump could be damaged. From a
bacteriological point of view, it becomes dangerous for the users.
Do not substitute this lter with any other material. Substitution may lead to contamination or
poor performance; use only Rossmax lter.
Removing condensation from the air tube
Follow the below procedures if there is moisture or medication remaining in the air tube used on
nebulizer kit/nasal washer.
1. Make sure that the air tube is connected to the air outlet on the compressor.
2. Remove the air tube from the nebulizer kit or nasal washer.
3. Switch on the device and pump air through the air tube to expel the moisture.
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the NK1000,
including cables specied by the manufacturer. Otherwise, degradation of the performance of
this device could result.
If higher IMMUNITY TEST LEVELS than those specied in Table 9 are used, the minimum separation
distance may be lowered. Lower minimum separation distances shall be calculated using the
equation specied in 8.10.
Manufacturers declaration-electromagnetic immunity
The NK1000 is intended for use in the electromagnetic environment specied below. The cus-
tomer or the user of the NK1000 should assure that is used in such and environment.
Immunity test Compliance levelIEC 60601 test level Electromagnetic environment-guid-
ance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the NK1000 includ-
ing cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz,
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sepa-
ration distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected
by absorption and reection from structures, objects and people.
More information on EMC compliance of the device can be obtained from Rossmax website:
www.rossmax.com
Accessories (Optional)
Name Model No Material
Disinfecting Method: Boiling
O: Applicable,
X: Not applicable
Nebulizer bottle set (N1
Bottle+Mouthpiece) NB_AC_001_00 PP O
Nebulizer bottle (N1) NB_AC_026_00 PP O
Mouthpiece ONB_AC_021_00 PP
Adult Mask XNB_AC_003_00 PVC
Child Mask XNB_AC_004_00 PVC
Air Tube (1.5M) NB_AC_005_00 PVC X
Air Filters (Ø15x10mm) x 5pcs NB_AC_006_00 PU X
Nasal Washer NW1 PP+ Silicone O
Nasal Aspirator NS1 PP+ Silicone O
Filter (nasal aspirator) - XNB_AC_032_00
Air tube (nasal aspirator) XNB_AC_033_00 PVC
* Please follow the disinfecting methods to boil and wash the accessories.
www.rossmax.com
Warranty Card
This instrument is covered by a 3 year guarantee from the purchase date. The guarantee is valid
only on presentation of the guarantee card completed by the dealer confirming purchase date
or the receipt. Nebulizer Accessories are not included. Opening or altering the instrument invali-
dates the guarantee. The guarantee does not cover damage, accidents or noncompliance with
the instruction manual. Please contact your local seller/dealer or www.rossmax.com.
Customer Name: ___________________________________________________________
Address: __________________________________________________________________
Telephone: ________________________________________________________________
E-mail address: _____________________________________________________________
Product Information
Date of purchase: ___________________________________________________________
Store where purchased:
___________________________________________________________________________
G
F-1
F-3
F-2
1
2
3
4
5
6
7
8
E-1 E-2 E-3
Model: NK1000
Nebulizer
WARNING: The symbol on this product means that its an electronic product and following
the European directive 2012/19/EU the electronic products have to be disposed on your lo-
cal recycling centre for safe treatment.


Produkt Specifikationer

Mærke: Rossmax
Kategori: Ikke kategoriseret
Model: NK1000

Har du brug for hjælp?

Hvis du har brug for hjælp til Rossmax NK1000 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Ikke kategoriseret Rossmax Manualer

Rossmax

Rossmax NK1000 Manual

1 September 2024

Ikke kategoriseret Manualer

Nyeste Ikke kategoriseret Manualer

Scarlett

Scarlett SC-HC63C107 Manual

9 November 2024
Alpine

Alpine APF-F291320VW Manual

9 November 2024
Klipsch

Klipsch CMS 4.1 Manual

9 November 2024
Alpine

Alpine MQR-III Manual

9 November 2024
Alpine

Alpine DDS-R69K Manual

9 November 2024
Alpine

Alpine KIT-8A3 Manual

9 November 2024
IFM

IFM CR1144 Manual

9 November 2024
IFM

IFM CR1143 Manual

9 November 2024
Auray

Auray SB-52SSB Manual

9 November 2024
Bluetti

Bluetti Charger 1 Manual

9 November 2024