Tektronix P6316 Manual


Læs nedenfor 📖 manual på dansk for Tektronix P6316 (6 sider) i kategorien niet gecategoriseerd. Denne guide var nyttig for 4 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/6
xx
Product Description
The P6316 digital probe connects the Tektronix MSO2000
Series of mixed-signal oscilloscopes to digital buses and signals
on your target system. The probe breaks out 16 data channels
over two 2x8-pin headers (GROUP 1 and GROUP 2). Each
header includes eight signals on one row and eight grounds
along the opposite row.
A pair of interchangable ying lead sets are included with the
probe. The lead sets bring out one ground connection on each
end and eight signal leads for connecting to individual test
points.
Connecting the Probe to the Oscilloscope
Connect the probe as shown in the illustration below.
1. Insert the probe label-side up into the receptacle.
2. To remove the probe, squeeze the buttons on the side and
pull out the probe.
Connecting the Probe to Your Circuit
Attach the probe to the circuit using the connectors and adapters
shown on the back of these instructions. Select the best method
for your needs, and then proceed to Setting up the Probe.
Setting up the Probe
Press the D15-D0 button to set and view the following
parameters of each digital channel:
Threshold voltage and vertical position
Signal height (set once for all 16 channels)
Channel label
The default settings are 1.4 V thresholds with no channel labels.
Use the B1 and B2 buttons to set and view bus characteristics
such as:
Clock type
Bus type (Serial or Parallel)
Bus width
Display format (Hex, Binary, ASCII)
Parallel bus setup information is resident on the MSO Family
of oscilloscopes. However, for other buses such as CAN and
I2C, you must have the appropriate application key. See your
oscilloscope manual or product data sheet for nomenclature
and ordering details.
Functional Check
Logic activity immediately displays on all connected, active
channels. If you do not see an active signal:
1. Press Trigger.
2. Select Edge for trigger type.
3. Select the channel that you are setting up as the source.
4. Press Autoset.
If you do not see an active signal, try another probe channel (or
analog probe) to verify circuit activity at the test point.
Specications
Table 1: Electrical and mechanical specications
Characteristic Description
Input channels 16
Threshold voltage range ±20 V
Threshold accuracy ±100 mV + 3% of threshold setting
Minimum signal swing 500 mV p-p
Maximum signal swing ±20 V centered around the
threshold voltage
Maximum nondestructive
input signal to probe
±40 V p-p
Input resistance 101 k
Input capacitance 8.0 pF
Minimum detectable input
pulse
5 ns
Probe length 0.9 m (3 ft)
Table 2: Environmental specications
Characteristics Description
Temperature
Maximum operating
Minimum operating
Nonoperating
+50 °C (+122 °F)
0 °C (+32 °F)
–40 °C to +71 °C
(–40 °F to +160 °F)
Humidity 5% to 95% relative humidity at up
to +30 °C (+86 °F)
5% to 60% RH above +30° C
(+86 °F) up to +50 °C (+122 °F),
non-condensing
Altitude
Operating
Nonoperating
3 km (9,843 ft) maximum
12 km (39,370 ft) maximum
Electrostatic Immunity The probe is not static sensitive
Equipment Recycling. This product complies
with the European Union’s requirements according
to Directive 2002/96/EC on waste electrical
and electronic equipment (WEEE). For more
information about recycling options, check the
Support/Service section of the Tektronix Web site
(www.tektronix.com).
Safety Summary
Connect and Disconnect Properly. Connect the probe output
to the measurement instrument before connecting the probe
to the circuit under test. Disconnect the probe input and the
probe ground from the circuit under test before disconnecting
the probe from the measurement instrument.
Observe All Terminal Ratings. To avoid re or shock hazard,
observe all ratings and markings on the product. Consult the
product manual for further ratings information before making
connections to the product.
Do not Operate Without Covers. Do not touch exposed
connections and components when power is present.
Avoid Exposed Circuitry. Do not touch exposed connections
and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is
damage to this product, have it inspected by qualied service
personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an
Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or
other property.
Symbols on the Product. These symbols may appear on the
product:
Typical Application
1. Use the P6316 probe to view digital signals on a system
bus.
2. Use analog probes, such as the P6139A or P2221 probes
to view analog waveform information.
P6316
Digital Probe
Instructions
x
1
*P071234500*
071-2345-00


Produkt Specifikationer

Mærke: Tektronix
Kategori: niet gecategoriseerd
Model: P6316

Har du brug for hjælp?

Hvis du har brug for hjælp til Tektronix P6316 stil et spørgsmål nedenfor, og andre brugere vil svare dig




niet gecategoriseerd Tektronix Manualer

Tektronix

Tektronix P6316 Manual

4 September 2024
Tektronix

Tektronix P6616 Manual

4 September 2024

niet gecategoriseerd Manualer

Nyeste niet gecategoriseerd Manualer

Pioneer

Pioneer CA-R-CG.001 Manual

18 September 2024
Pioneer

Pioneer CA-R-MAC.001 Manual

18 September 2024
Pioneer

Pioneer CA-R-PI.187 Manual

18 September 2024
NEC

NEC LED-FE019i3 Manual

18 September 2024
Exquisit

Exquisit EKS180-4-060E Manual

18 September 2024
Smeg

Smeg SA661X Manual

18 September 2024
Pioneer

Pioneer CA-R-TOY.003AE Manual

18 September 2024
Pioneer

Pioneer CA-R-LA.001 Manual

18 September 2024
SereneLife

SereneLife SLTLWS550 Manual

18 September 2024
SereneLife

SereneLife SLTLWS650 Manual

18 September 2024