PN 2094740 May 2003 ©2003 Fluke Corporation. All rights reserved. Printed in U.S.A.
700PMP
Precision Pressure Pump
Instruction Sheet
Introduction
The Fluke 700PMP is a tube type, cylindrical
shaped hand pump with a "T" handle at the
compressor end and a round knob at the volume
adjust end. The overall length is between 11.5"
(full extension) and 8.5" (fully collapsed). The
pump incorporates a needle valve for venting
and a volume adjust vernier for precision
adjustment of pressure.
Pressure connections are made through one 1/8"
NPT internally threaded fitting. The pump is
small in size, lightweight and ruggedly
constructed to withstand typical field use.
Operating Instructions
Producing Pressure
1. Connect the pump's port to the instrument to
be calibrated or checked. Use small-diameter
tubing as short in length as possible (this will
maximize the pressure adjustment range).
2. Set the FINE ADJUST knob to the full
counterclockwise position.
3. Turn the bleed valve knob fully
counterclockwise to relieve all system
pressure and zero any measuring devices.
4. Turn the bleed valve knob fully clockwise to
close.
5. Repeatedly move the "T" handle in and out to
generate the desired pressure.
6. Use the FINE ADJUST knob to bring up the
pressure to the precise level.
7. Use the BLEED VALVE to lower the pressure
from the pressure generated. Opening the
BLEED VALVE 1/4 turn will lower the
pressure very gradually. Opening it 1/2 turn
will release the pressure faster and opening it
3/4 turn will quickly and safely release all the
pressure in the system.
9Warning
It is imperative that all system
pressure is relieved prior to making
any connections or disconnections.
Failure to relieve system pressure
could result in serious personal
injury or equipment damage. Even
nominal pressure values can
generate extreme force if fitting or
tubing failure occurs due to improper
installation or usage. Since the pump
is capable of generating pressures
exceeding 100 psig/70 bar, it is
important that all pressure
connections and test procedures be
done by qualified service personnel,
according to standard engineering
practices, to prevent possible
personal injury or equipment
damage.
Connections
To install a pressure fitting in the pump:
1. Turn the BLEED VALVE counterclockwise to
bleed any pressure.
2. Use a 5/8" open-end wrench on the input port
to prevent it from rotating while tightening the
supply fitting with a 5/8" open-end wrench.
Leak Prevention and Detection
In order to obtain maximum pressure indication
stability, leaks must be avoided. It is strongly
recommended that either Teflon® tape or
commercial pipe sealant is used at all tapered
fittings and connections. If Teflon® tape is used,
care must be taken that the proper amount is
applied. Excessive tape may fray and cause
plugging of relief valves orifices, nozzles, etc.
Overuse of pipe sealant may cause similar
problems.


Produkt Specifikationer

Mærke: Fluke
Kategori: Pomp
Model: 700 PMP

Har du brug for hjælp?

Hvis du har brug for hjælp til Fluke 700 PMP stil et spørgsmål nedenfor, og andre brugere vil svare dig




Pomp Fluke Manualer

Fluke

Fluke 700HTH-1 Manual

22 August 2024
Fluke

Fluke 730HTP70M Manual

14 August 2024
Fluke

Fluke 730PTP10M Manual

23 Juli 2024
Fluke

Fluke 730LTP40K Manual

23 Juli 2024
Fluke

Fluke 700 PMP Manual

22 Juli 2024

Pomp Manualer

Nyeste Pomp Manualer

AstralPool

AstralPool Eco Range Manual

15 December 2024
Gardenline

Gardenline GLRP 300 Manual

15 December 2024
Ozito

Ozito PTP-012A Manual

15 Oktober 2024
MSW

MSW -MN34 Manual

15 Oktober 2024
MSW

MSW -MN34M Manual

15 Oktober 2024
MSW

MSW -AOP40S Manual

15 Oktober 2024
Nefit

Nefit VentiLine Manual

12 Oktober 2024