Microlife MT 710 Manual


Læs nedenfor 📖 manual på dansk for Microlife MT 710 (1 sider) i kategorien Termometer. Denne guide var nyttig for 32 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/1
MT 710
IB MT 710 EN 1220 - Revision Date: 2020-05-15
4 2 1 35
Description of this Thermometer
Important Safety Instructions
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device and
keep for future reference.
This device is only to be used for measuring human body tempera-
ture through oral, rectal or axillary. Do not attempt to take tempera-
tures at other sites, such as in the ear, as it may result in false read-
ings and may lead to injury.
Do not use this device if you think it is damaged or notice anything
unusual.
We recommend cleaning this device according to the cleaning
instructions before first use for personal hygiene.
The minimum measurement time until the beep is heard must be
maintained without exception!
Consider that different measurement locations may require
continued measuring even after the beep, see section «Measuring
methods / Normal body temperatur.
Do not attempt rectal measurements on persons with rectal disor-
ders. Doing so may aggravate or worsen the disorder.
Ensure that children do not use this device unsupervised; some
parts are small enough to be swallowed. Be aware of the risk of
strangulation in case this device is supplied with cables or tubes.
Do not use this device close to strong electromagnetic fields such as
mobile telephones or radio installations. Keep a minimum distance
of 3.3 m from such devices when using this device.
Protect the device from impact and dropping!
Avoid bending the thermometer probe more than 45°!
Avoid ambient temperatures above 60 °C. NEVER boil this device!
Use only the commercial disinfectants listed in the section «Cleaning
and Disinfecting» to clean the device to avoid damage to the device.
We recommend this device is tested for accuracy every two years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test.
WARNING: The measurement result given by this device is
not a diagnosis! Do not rely on the measurement result only.
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Read the instructions carefully before using this device.
Type BF applied part
Turning on the Thermometer
To turn on the thermometer, press the ON/OFF button 1; a short
beep signals «thermometer ON». A display test is performed. All
segments should be displayed.
Then «-- -» and a flashing «°C» appear at the display field 2. The
thermometer is now ready for use.
Function Test
Correct functioning of the thermometer is tested automatically each
time it is turned on. If a malfunction is detected (measurement inaccu-
racy), this is indicated by «ERR» on the display, and a measurement
becomes impossible. In this case, the thermometer must be replaced.
Using the Thermometer
Choose the preferred measuring method. When taking a measure-
ment, the current temperature is continuously displayed and the «°C»
symbol flashes. If the beep is heard 10 times and the «°C» is no longer
flashing, this means that the measured increase in temperature is less
than 0.1 °C in 6 seconds.
To prolong the battery life, turn off the thermometer by briefly pressing
the ON/OFF button 1. Otherwise the thermometer will automatically
turn off after about 10 minutes.
Reviewing the last Measurement
Press ON/OFF button
1
for
2
seconds, when turning on the thermom-
eter. The last measurement will be displayed together with a
« »M
for
memory. After
2
seconds the thermometer will automatically switch off.
Measuring methods / Normal body temperature
In the armpit (axillary) / 34.7 - 37.3 °C
Wipe the underarm with a dry towel. Place the measuring sensor 4
under the arm into the center of the armpit so the tip is touching the
skin and position the patient’s arm next to the patient’s body. This
ensures that the room air does not affect the reading. Because the
axillary takes more time to reach its stable temperature wait at least
5 minutes, regardless of the beep sound.
In the mouth (oral) / 35.5 - 37.5 °C
Do not eat or drink anything hot or cold 10 minutes before the
measurement. The mouth should remain closed up to 2 minutes
before starting a reading.
Position the thermometer in one of the two pockets under the
tongue, to the left or right of the root of the tongue. The measuring
sensor 4 must be in good contact with the tissue. Close your mouth
and breathe evenly through the nose to prevent the measurement
from being influenced by inhaled/ exhaled air.
If breathing through the mouth cannot be prevented due to blocked
airways, use another measuring method.
Approx. measuring time: 30 seconds!
In the anus (rectal) / 36.6 - 38.0 °C
Carefully insert the measuring sensor of the thermometer 2 to 4
3 cm into the anal aperture.
The use of a probe cover and the use of a lubricant is recommended.
If you are unsure of this measurement method, you should consult a
professional for guidance/training.
Approx. measuring time: 30 seconds!
Cleaning and Disinfecting
For disinfection in home use environment, use a 70% Isopropyl
alcohol swab, or a cotton tissue moistened with 70% Isopropyl alcohol
to wipe surface pollutants off the thermometer probe (note: consider
the application and safety instruction of the disinfectant manufacturer).
Always start wiping from the end of the thermometer probe (approx. at
the middle of the thermometer) towards the thermometer tip. After-
wards the entire thermometer probe (see number 5 in the drawing)
should be immersed in 70% Isopropyl alcohol for at least 5 minutes
(max. 24 hours). After immersion, let the disinfectant dry off for 1
minute before next use. Avoid immersing or wiping the display to
protect it from fading.
The thermometer is not intended for professional use.
Battery Replacement
When the «» symbol (upside-down triangle) appears in the display,
the battery is flat and needs replacing. To replace the battery remove
the battery compartment cover 3 from the thermometer. Insert the
new battery with the + at the top. Make sure you have a battery of the
same type to hand. Batteries can be purchased at any electrical store.
1 ON/OFF button
2 Display
3 Battery compartment cover
4 Measuring sensor / measuring tip
5 Cleaning and disinfecting area (thermometer probe only)
EN
Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Guarantee
This device is covered by a lifetime guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife will
repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance with the
instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Battery.
Should guarantee service be required, please contact the dealer from
where the product was purchased, or your local Microlife service. You
may contact your local Microlife service through our website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will
be granted if the complete product is returned with the original invoice.
Repair or replacement within guarantee does not prolong or renew the
guarantee period. The legal claims and rights of consumers are not
limited by this guarantee.
Type: Maximum thermometer
Measurement range: 32.0 °C to 42.9 °C
Temp. < 32.0 °C: display «-- -» for low (too low)
Temp. > 42.9 °C: display «H» for high (too high)
Measurement accuracy: ± 0.1 °C; 34 °C - 42 °C
± 0.2 °C; 32.0 - 33.9 °C and 42.1 - 42.9 °C
Operating conditions: 10 - 40 °C; 15-95 % relative maximum humidity
Storage conditions: -25 - +60 °C; 15-95 % relative maximum humidity
Battery: LR41 (1.5V) / SR41 (1.55V)
Battery lifetime: approx. 4500 measurements (using a new battery)
IP Class: IP22
Reference to standards: EN 12470-3, clinical thermometers;
ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC);
IEC 60601-1-11
Expected service life: 5 years or 10000 measurements
Microlife AG, Espenstrasse 139
9443 Widnau / Switzerland


Produkt Specifikationer

Mærke: Microlife
Kategori: Termometer
Model: MT 710

Har du brug for hjælp?

Hvis du har brug for hjælp til Microlife MT 710 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Termometer Microlife Manualer

Termometer Manualer

Nyeste Termometer Manualer