Kyoritsu KEW 2433R Manual

Kyoritsu Multimeter KEW 2433R

Læs nedenfor 📖 manual på dansk for Kyoritsu KEW 2433R (2 sider) i kategorien Multimeter. Denne guide var nyttig for 3 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/2
INSTRUCTIONMANUAL
KEW SNAP
 
2433R
KEW SNAP Series
LEAKAGECURRENTTESTER
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
1. SAFETY WARNINGS
This instrument has been designed and tested according to IEC
Publication
61010
: Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety
ruleswhichmustbeobserved bytheusertoensuresafeoperation
ofthe instrument and to retain it in safe condition.  Therefore, read
through these operating instructions before starting using the
instrument.
# WARNING
Readthrough and understand instructions contained in this
manualbeforestartingusingtheinstrument.
Save and keep the manual handy to enable quick reference
whenevernecessary.
Theinstrumentistobeusedonlyinitsintendedapplications.
Understandandfollowall thesafetyinstructionscontainedin
themanual.
Failuretofollowtheinstructionsmaycauseinjury,instrument
damage and/or damage to equipment under test. Kyoritsu
is by nomeans liable for any damage resultingfrom the
instrumentincontradictiontothiscautionarynote.
The symbol
#
indicatedon the instrumentmeans that the user
mustrefertorelated partsofthemanualforsafeoperationofthe
instrument. Besureto carefully readthe instructionsfollowing
each
#
symbolinthismanual.
#DANGER
isreservedforconditionsandactionsthatarelikely
tocauseseriousorfatalinjury.
#WARNING
 is reserved for conditions and actions that can
causeseriousorfatalInjury.
#
CAUTION
isreservedforconditionsandactionsthatcancause
minorinjuryorInstrumentdamage.
Thefollowingsymbolsareusedandmarkedontheinstrumentand
intheinstructionmanual.Pleasecarefullycheckbeforestartingto
usetheinstrument.
#
Refertotheinstructionsinthemanual.
This symbol is marked where the user must refer to the
instruction manual so as not to cause personal injury or
instrumentdamage.
Indicates an instrument with double or reinforced
insulation.
Indicates that this instrument can clamp on bare
conductors when measuring avoltage corresponding to
the applicable Measurement category,which is marked
nexttothissymbol.
IndicatesAC(AlternatingCurrent).
#
DANGER
Never make measurement on a circuit having potential of
300
VACorgreater.
Do not attempt to make measurement in the presence of
flammablegasses.Otherwise,theuseoftheinstrumentmay
causesparking,whichleadstoanexplosion.
Thetransformerjawsaremadeofmetalandtheirtipsarenot
completelyinsulated.Beespeciallycarefulaboutthepossible
shorting where the equipment under test has exposed metal
parts.
Neverattempttousetheinstrumentifitssurfaceoryourhand
is
wet.
Do not exce ed the maximum allo wable input of any
measurementrange.
Never open the battery compartment cover when making
measurement.
Neverfry to make measurement if any abnomal conditions.
suchasbrokenTransformerjawsorcaseisnoted.
Theinstrument is be used only in its intended applicationsor
conditions. Otherwise, safety functions equipped with the
instrument doesn
twork, andinstrument damage orserious
personalinjurymaybecaused.
Keepyourfingersandhandsbehindtheprotectivefingerguard
duringmeasurement.
#
WARNING
Never attempt to make any measurement, if any abnormal
conditionsarenoted,suchasbrokencase,crackedtestleads
andexposedmetalparts.
Do notinstall substituteparts or makeany modification to
the instrument.  Return the instrument to Kyoritsu or your
distributorforrepairorre-calibration.
Do not try to replace the batteries if the surface of the
instrumentiswet.
Always switch off the instrument before opening the battery
compartmentcoverforbatteryreplacement.
#
CAUTION
Make sure that the range selector switchis set to an
appropriatepositionbeforemakingmeasurement.
Do not expose the instrument to the directsun, extreme
temperaturesordewfall.
Besuretosettherangeselectorswitchtothe"OFF"position
afteruse. When theinstrument will not bein use for along
periodoftime,placeitinstorageafterremovingthebatteries.
Useadampcloth anddetergentforcleaning the instrument.
Donotuseabrasivesorsolvents.
MeasurementCategory:
Toensuresafeoperationofmeasuringinstruments,IEC
61010
establishessafetystandardsforvariouselectricalenvironments,
categorized as O to CAT IV, and called measurement
categories.Higher-numberedcategoriescorrespondtoelectrical
environments with greater momentary energy, so a measuring
instrument designed for CAT III environments can endure
greatermomentaryenergythanonedesignedforCATII.
O :Circuits whichare not directlyconnected to the mains
powersupply.
CATII :Electrical circuits of equipment connected to an AC
electricaloutletbyapowercord.
CATIII :Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distributionpaneltooutlets.
CATIV:The circuit from the service drop to the service
entrance, and to the power meter and primary over-
currentprotectiondevice(distributionpanel).
Incoming wire
: Device which is
not directly
connected to the
mains power supply
Socket
Interior wiring
2. FEATURES
DigitalclamptesterforACleakagemeasurement.
Accuratetrue-RMSreadingofACcurrentwithdistortedwaveform.
Least affected by external magnetic field, providing wide
measuringrangefromverysmalltolargecurrents.
Designed to safety standard IEC
61010
-
2
-
032
: Measurement
categoryCAT
,
300
Vandpollutiondegree
2
.
Teardrop shapedjaws for ease of use in crowded cable areas
andothertightplaces.
Dataholdfunctiontoallowforeasyreadingsindimlylitorhard-to-
reachlocations.
Providesfilteringfunctiontoremovehighfrequencygeneratedby
suchequipmentasinverters.
Peakhold function toallow for measurementof current variation
asshortas
10
msec.
Auto-power-offfunctionpreventsunnecessarypowerconsumption
Dynamicrangeof
4200
countsfullscale.
Largeeasy-to-readLCDdisplaywithletterheightof
13
mm.
Operationconfirmingbeeps.
Insulationbarrieratthetipoftransformerjawsforimprovedsafety.
3. SPECIFICATIONS
Measuringrangesandaccuracy(Sinewave)
Range Resolution MeasuringRange Accuracy(Frequencyrange)
40
mA
0
.
01
mA
0
40
.
00
mA
0〜100A
 ±
1
.
0
%rdg
±
5
dgt (
50
/
60
Hz)
 ±
2
.
5
%rdg
±
10
dgt (
20
Hz
1
kHz)
100〜300A
 ±
1
.
0
%rdg
±
5
dgt (
50
/
60
Hz)
 ±
2
.
5
%rdg
±
10
dgt (
40
Hz
1
kHz)
300〜400A
 ±
2
.
0
%rdg (
50
/
60
Hz)
 ±
5
.
0
%rdg (
40
Hz
1
kHz)
400
mA
0
.
1
mA
0
400
.
0
mA
400
A
0
.
1
A
0
400
.
0
A
CF(Crestfactor)
3
(
45
65
Hz,lessthan
600
APeak)
    
100
400
A
sinewave
2
%rdg
Countsequaltoorlessthan
3
countsarecorrectedtozero
Accuracy-insuredFrequencyrangeof
50
/
60
Hzmodeis
50
/
60
Hz.
The max indication at the
40
mA/
400
mA range is
6000
 counts.
Minute current may existwhilezerois displayed at
400
A/
400
mA
range.Measurementshouldbemadealsoatalowerrange.
Conversion method:
Rmsvaluedetection
Operating System:
 Sequentialcomparison
Display:
  LCDwithmax.readingof
4200
(
400
Arange),
6000
(
40
/
400
mArange)
Low battery warning:
 "BATT"markappearsonthedisplay
Overrange Indication:
 "OL"appearsonthedisplaywhenupperlimit
ofmeasuringrangeisexceeded
Response Time:
 Approx.
2
seconds
Sample Rate:
 Approx.
2
.
5
timespersecond
Accuracy-insured
 
23
℃±
5
,relativehumidity
85
%orless
Temperature and
 (withoutcondensation)
Humidity Ranges:
Operating Temperature
 
0
-
40
,relativehumidity
85
%orless
and Humidity Ranges:
 (withoutcondensation)
Storage Temperature
 -
20
-
60
,relativehumidity
85
%orless
and Humidity Ranges:
 (withoutcondensation)
Operable altitude:
2000
morlessabovesealevel(indooruse)
Power Source:
 Two
1
.
5
VR
03
(AAA)batteries
Current Consumption:
 Approx.
21
mA
Measurement Time:
 Approx.
24
hours
Auto-power-off Function:
Turns power off about
10
minutes afterthe
lastswitchoperation
Safety Standard:
 IEC
61010
-
1
  IEC
61010
-
2
-
032
  MeasurementCAT
300
V,pollutiondegree
2
EMC directive:
 EN
61326
-
01
,EN
61326
-
2
-
2
Environmetal standard:
 EURoHSdirectivecompliant
Overload Protection:
480
AACmax.for
10
seconds
Withstand Voltage:
 
4240
VACrms (
50
/
60
Hz) for
5
 sec. between
metal part of transformer jaws and housing
case(excepttransformerjawcase)
Insulation Resistance:
 
50
M
Ω
 or greater at
1000
V between metal
part of transformer jaws and housing case
(excepttransformerjawcase)
Conductor Size:
 Approx.
40
mmindiametermax.
Dimensions:

185
(L)
×
81
(W)
×
32
(D)mm
Weight:
  Approx.
270
gincludingbatteries
Accessories:
 TwoR
03
(AAA)batteries
  CarryingcaseModel
9052
  Instructionmanual
Reference
*Effective Value (RMS)
Most alternating currents and
voltag es are expre s s ed in
effective values, which are
alsoreferredtoasRMS(Root-
Mean-Square)values.
The e ffective value is the
square root of the average of
squareofalternatingcurrentor
voltagevalues.Manyclampmetersusingaconventionalrectifyingcircuit
have"RMS"scalesforACmeasurement.Thescalesare,however,actually
calibratedintermsoftheeffectivevalueofasine wavethoughtheclamp
meteris respondingto theaverage value.The calibrationis done with a
conversion factor of
1
.
111
 for sine wave, which is found by dividing the
effective value by the average value. These instruments are therefore in
erroriftheinputvoltageorcurrenthassomeothershapethansinewave.
*CF (Crest Factor) is found by dividing the peak value by the effective
value.
Examples:
  
Sinewave:CF=
1
.
414
Squarewavewitha
1
:
9
dutyratio:CF=
3
4. INSTRUMENT LAYOUT
●LCD
Low Battery Warning
Frequency Response :
Wide
Frequency Response :
50/60Hz
Data Hold Indication
Peak Hold Indication
Unit
Di s pl ay
Frequency Selector Button
Peak Hold Button
Data Hold Button
Tr a ns fo rm J aw s
Ba r ri er Tr i gg er
Ra n ge S el ec to r Sw it ch
Ha n d St ra p
Barrier:It is a part providing protection against electrical shock and
ensuringtheminimumrequiredairandcreepagedistances.
5. PREPARATIONS FOR MEASUREMENT
5
-
1
CheckingBatteryVoltage
Setthe RangeSelector Switchto anyposition otherthan theOFF
position.Ifthemarksonthedisplayisclearlyvisiblewithout"BATT"
markshowing,batteryvoltageisOK.Ifthedisplayblanksor"BATT"
is indicated, replace the batteries according to section
8
: Battery
Replacement.
NOTE
When the instrument is left poweredon, the auto-power-off
functionautomatically shut the power off; Thedisplay blanks
even ifthe Range Selector Switchis set toa position other than
the OFF position in this state.  To poweron the instrument, turn
the Range Selector Switch or press the Data Hold Button. If the
display still blanks, thebatteriesarecompletelyexhausted.
Replacethebatteries.
5
-
2
CheckingSwitchSetting
MakesurethattheRange SelectorSwitchis settothe appropriate
range.
Al s o ma ke s ur e t hat  da ta  ho ld  fu nc ti on i s n ot  en ab le d.  If
inappropriate range is selected, desired measurement cannot be
made.
6. OPERATING INSTRUCTIONS
6
-
1
CurrentMeasurement
# DANGER
Inordertoavoidpossibleshockhazard,nevermakemeasurement
oncircuitshavingapotentialof
300
VACorgreater.
The transformer jaws are made of metal and theirtips are not
completelyinsulated.Beespecially carefulaboutthe possible
shortingwheretheequipmentundertesthasexposedmetalparts.
Never makemeasurement withthe batterycompartment cover
removed.
Whenmeasuringcurrentis
300
Aormore(
400
Hzormore),besure
tostopmeasurementwithin
5
minutes.Otherwise,transformer jaws
mayheattocauseafireordeformationofmoldedparts,whichwill
degradeinsulation.
Keepyourfingersandhandsbehindthebarrierduringmeasurement.
# CAUTION
Take sufficient care to not to apply shock, vibration or
excessiveforcetothejawtips.Otherwise,preciselyadjusted
TransformerJawtipswillbedamaged.
When a foreign substance is stuck in the jaw tips or they
cannot properly engage, the transformer jaws do not fully
close.Insuchacase,donotreleasethejawtriggerabruptly
orattemptto closethe transformer jaws byapplying external
force.  Make sure that the jaws close by themselves after
removingtheforeignsubstanceormakingthemfreetomove.
The maximum size of a conductor to be tested is
40
mm
in diameter.  Accurate measurement cannot be made on a
conductor larger than this, because the transformer jaws
cannotfullyclose.
Whenmeasuringlargecurrent,thetransformerjawsmaybuzz.
Thishasnoeffectontheinstrument'sperformanceorsafety.
SensitivetransformerjawsareusedforLeakageclampmeter.
Because of the characteristics of transformer jaws,which
canbe opened and closed, it is impossible to eliminate the
interference of external magnetic field completely. If there
are something,which generatinglarge magnetic field,at a
nearby site, current value can be displayed (
0
cannot
be displayed.) before clamping on the conductor. For such
a case, please use the instrument at a location far from the
thing,whichgeneratingmagneticfield.
Followingarethetypicalthingsgeneratingmagneticfield.
  ・
Conductorfedlargecurrent
  ・
Motor
  ・
Equipmentwhichhasmagnet
  ・
Integratingwattmeter
(
1
)SettheRangeSelectorSwitchtothedesiredposition.Currentto
measureshouldbewithintheselectedmeasuringrange.
(
2
)Normalmeasurement(SeeFig.
1
,
2
):
Pressthejawtriggertoopenthetransformerjawsandclosethem
overoneconductoronly.Measuredcurrentvalueisshownonthe
display.Earthleakagecurrentorsmallcurrentthatflowsthrough
agroundedwirecanalsobemeasuredbythismethod.
(
3
)Measuringoutofbalanceleakagecurrent(SeeFig.
3
):
Clamp onto allconductors except a grounded wire. Measured
currentvalueisshownonthedisplay.
Fig.
1
Loadcurrent
       
Fig.
Earthleakagecurrent
Load
Load
Fig.
3
Measuringoutofbalanceleakagecurrent
Load
Load
3
-phase
3
-wiresystem
(In
4
-wiresystem with neutral,
clampontoall
4
wires)
Single-phase
3
-wiresystem
(in
3
-wiresystem with neutral,
clampontoall
3
wires)
6
-
2
HowtoUseFrequencySelectorButton
Whenhighfrequenciesfromsuchequipmentasinvertersarepresent
inthecircuitundertest,theinstrumentmeasuresACcurrentofnot
only
50
Hzor
60
Hzoffundamentalfrequencybutalso ofthese high
frequenciesandharmonics.
To eliminate the effect of such high frequency noise and measure
AC currentof
50
Hz or
60
Hzfundamental frequency,a "high-cut"
filtercircuit in incorporated into the instrument which works when
"
50
/
60
Hz" frequency response is selected with the Frequency
Selector Button. Cut-off frequency of the "high-cut" filter is about
160
Hzwithattenuationcharacteristicofapprox.-
24
dB/octave.
WhentheFrequencySelectorButtonispressed,"
50
/
60
Hz"markis
shownontheleftsideofthedisplay.WhentheFrequencySelector
Buttonis pressedagain, fequency responseis switchedto WIDE
with"WIDE"markshownonthedisplay.
OutputcharacteristicareshowninFig.
4
.
Fig.
4
KEWSNAP
2433
RFrequencyCharacteristic
-30
-20
-10
0
10
Attenuation dB〔 〕
1 5k k50 500 10k 50k20k22001002010 100k
100%
50%
20%
10%
5%
Percentage
Typical characteristic "WIDE" frequency response
     ,fc=8kHz -24dB/cot
Typical characteristic "50/60Hz"
frequency response
     ,fc=160Hz -24dB/cot
Frequency Hz〔 〕
Note:
Characteristic of -
24
dB/octiave means that signal magnitude
declinestoaboutonesixteenthofthatattheinitialfrequencywhen
frequency doubles. KEW SNAP
2433
R have the following two
settingsfortheFrequencySelectorButton.
WIDE(
20
Hz-approx.
kHz):
Permits measurement of currents of
fundamental frequencies as well as
currentsofhighfrequenciesgenerated
bysuchequipmentasinverters
50
/
60
Hz(
20
-approx.
160
Hz):Filters out high frequency currents
and measures current of fundamental
frequencyonly
Recently there has been increased use of power through inverters,
switchingregulators, etc. When the high frequency noise from such
appliances leaks or flowsinto the ground through capacitors not
filtering completely, the earth leakagebreaker may trip even though
thereisno"actual"leakage.Insuchacase,theinstrumentdonotgive
leakagecurrentreadingif"
50
/
60
Hz"frequencyresponseisselected.
Take current readings with the
50
/
60
Hz and WIDE frequency
responses respectively to make effective use of the Frequency
SelectorButton.
6
-
3
PeakCurrentMeasurement
(
1
)SettheRangeSelectorSwitchtothedesiredposition.(Currentto
measureshouldnotexceedtheselectedmeasuringrange.)
(
2
)Select"WIDE"or"
50
/
60
Hz"withtheFrequencySelectorButton.
(
3
)With the transformer jaws clamped onto the conductor under
test,pressthePeakHoldButtontosettheintermenttothepeak
measurementmode.("P"isshownonthedisplay.)
(
4
)The display reads
1
/
2
 of thepeak current value. Therefore,an
rms reading is shown when current of a sinusoidal waveform is
measured.
(
5
)Afterpeakmeasurement,pressthe PeakHoldButtontoreturnto
thenormalmeasurementmode.
Note:When leakage current is measured in the peak measurement
mode, the reading may change if the transformer jaws are
openedand closed.Please readthedisplay with theconductor
under test clamped, otherwise, after fixing the display by using
the data hold function, please remove the instrument from the
conductor to be measured, and read the display. To measure
thepeakcurrentagain,pleasereleasethedatahold,andreturn
theinstrument to the normal measurement modeonce with the
PeakHoldButton,thensetitinthepeakmeasurementmode.
Countsequaltoorlessthan
5
countsarecorrectedtozero.
7. OTHER FUNCTIONS
7
-
1
Auto-Power-OffFunction
Thisisafunctiontopreventtheinstrumentfrombeingleftpowered
on and conserve battery power. The instrument automatically turns
offabout
10
 minutesafter the last switch or button operation.To
returntothenormalmode,turntheRangeSelectorSwitchtoOFF,
thentothedesiredposition.
DisablingAuto-Power-OffFunction:
Todisabletheauto-power-offfunction,powerontheinstrumentwith
theData Hold Button pressed. About
3
seconds after powering on
theinstrument,"P.OFF"isshownonthedisplay.Toenabletheauto-
power-offfunction,turnonthe instrumentwithoutpressingtheData
HoldButton.
Note: The auto-power-off function is disabled in the peak
measurementmode.
7
-
2
DateHoldFunction
Thisisafunctiontofreezethereadingsonthedisplay. Whenthe
DataHoldButtonispressed once,thecurrentreadingisheld even
though current under test varies. "H" mark is shown on the upper
rightcornerofthedisplay.
Toexitthedataholdmode,presstheDataHoldButtonagain.
Note:Whentheauto-power-offfunctionworkswhiletheinstrumentis
inthedataholdmode,dataholdiscancelled.
92
-
1557
H
5
-
22
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs
describedinthismanualwithoutnoticeandwithoutobligations.
8. BATTERY REPLACEMENT
# WARNING
Inorder toavoid possibleshock hazard,always setthe Range
SelectorSwitchtotheOFFpositionbeforetryingtoreplacethe
batteries.
# CAUTION
Donotmixnewandoldbatteries.
Installbatteriesintheorientationasshowninsidethebattery
compartment,observingcorrectpolarity.
Whenthe battery voltage warning mark "BATT" is shown on the
top left corner of the LCD, replace the batteries.  Note that the
displayblanksand"BATT"markisnotshownifthebatteriesare
completelyexhausted.
(
1
)SettheRangeSelectorSwitchto"OFF."
(
2
)Loosenthebattery-compartment-cover-fixingscrewonthelower
backoftheinstrument.
(
3
)Replace the batteries with two new R
03
 (AAA)
1
.
5
V
batteries.
(
4
)Putthebatterycompartmentcoverbackinplaceandtighten
thescrew.
Note:Foruseforalongperiodoftime,usealkalinebatteries(LR
03
).
Screw Battery
Compartment
Cover
Batteries
DisposetheBatteries
Thismarkingmeanstheyshallbesortedoutandcollectedas
ordainedinDIRECTIVE
2006
/
66
/EC.
This directive is valid only in the EU. When you remove
batteries from this product and dispose them, discard them in
accordancewith domestic law concerning disposal. Take a right
actiononwastebatteries,becausethecollectionsystemintheEU
onwastebatteriesareregulated.
WasteElectricalandElectronicEquipment(WEEEE)
Thismarkingmeanstheyshallbesortedoutandcollectedas
ordainedinDIRECTIVE
2006
/
66
/EC.
This directive is valid only in the EU. When you remove
batteries from this product and dispose them,discard them
inaccordance with domestic lawconcerningdisposal. Take aright
actiononwastebatteries,becausethecollectionsystemintheEU
onwastebatteriesareregulated.


Produkt Specifikationer

Mærke: Kyoritsu
Kategori: Multimeter
Model: KEW 2433R

Har du brug for hjælp?

Hvis du har brug for hjælp til Kyoritsu KEW 2433R stil et spørgsmål nedenfor, og andre brugere vil svare dig




Multimeter Kyoritsu Manualer

Kyoritsu

Kyoritsu KEW 2433R Manual

16 November 2024
Kyoritsu

Kyoritsu KEW 2432 Manual

16 November 2024
Kyoritsu

Kyoritsu KEW2046R Manual

20 September 2024
Kyoritsu

Kyoritsu KEW 2510 Manual

29 August 2024
Kyoritsu

Kyoritsu KEW 2500 Manual

29 August 2024
Kyoritsu

Kyoritsu KEW 2413R Manual

29 August 2024
Kyoritsu

Kyoritsu KT 200 Manual

29 August 2024
Kyoritsu

Kyoritsu 2431 Manual

29 August 2024
Kyoritsu

Kyoritsu 2434 Manual

28 August 2024

Multimeter Manualer

Nyeste Multimeter Manualer