Sanwa DLC470 Manual


Læs nedenfor 📖 manual på dansk for Sanwa DLC470 (3 sider) i kategorien Måleudstyr. Denne guide var nyttig for 7 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere

Side 1/3
− 1 − − 2 − − 3 − − 4 − − 5 −
[1] SAFETY PRECAUTIONS
*Before use, read the following safety
precautions.
This instruction manual explains how to use
your digital leakage clamp meter DLC470.
Before use, please read this manual thoroughly
to ensure correct and safe use. After reading it,
keep it together with the product for reference
to it when necessary.
Using the product in a manner not specified in
this manual may cause damage to the protection
function of the product.
The instructions given under the headings of
WARNING and CAUTION must be followed
to prevent accidental and burn electric shock.
1-1 Explanation of Warning Symbols
The meanings of the symbols used in this
manual and attached to the product are as
follows:
: Very important instructions for safe use.
The warning messages are intended to
prevent accidents to operating personnel
such as burn and electric shock.
The caution messages are intended to
prevent incorrect handling which may
damage the product.
: High voltage hazard : Ground
: Resistance : Buzzer
: Backlight
: Double insulation or reinforced insulation
1-2 Warning Messages for Safe Use
The following instructions are intended to
prevent personal injury such as burn and
electric shock.
Be sure to follow them when using the meter.
1. Never use the meter for the power lines
exceeding 600 Vrms to ground.
2. Voltages over AC 30 Vrms (42.4 Vpeak)
or DC 60 V are hazardous to human
body. Take care so as not to touch them.
3. Never input signals exceeding the
maximum rated input value (see 1-3).
4. Never use the meter near equipment
which generates strong electromagnetic
waves or is charged.
5. Never use the meter if the meter or test
leads are damaged or broken.
6. Never use the meter with the case or
battery lid removed.
7. During measurement, keep your fingers
behind the finger guard of test leads
and the meter of the test leads.
8. During measurement, do not change
function switch of the meter nor replace
the plugs to other terminals.
9. Before starting measurement, make
sure that the function and range are
properly set.
10. Never use the meter when it is wet or
with wet hands.
11. Be sure to use the specified type of test
leads.
12. Never attempt repair or modification,
except for battery replacement.
13. Always conduct start-up inspection and
check the meter at least once a year.
14. This meter is for indoor use only.
1-3 Overload Protection
The maximum rated input value and overload
protection have been established for the input
terminals of each function.
Function Input
Terminal
Max. Rated
Input Value
Max. Overload
Protection
DCVACV
and
DC/AC
600 V DC/AC
660 V
Do not input
voltage
ACmA CT AC 600 mA AC 100 A
ACA AC 400 A AC 450 A
[2] APPLICATIONS AND FEATURES
2-1 Applications
This is a digital leakage clamp meter designed
for measurement in the ranges of CAT. III 600 V.
This meter is useful for measuring leakage or
load current on power lines and equipment.
2-2 Features
Compatible to IEC61010 CAT.III
Extract frequency range of mainly ACmA 50
Hz / 60 Hz and ACA, with band-pass filter
functions.
Measures mA with resolution 0.01 mA
MAX MIN hold and data-holding functions
Backlight function
Auto Power Off function
Measurement categories (Overvoltage
categories)
CAT. II : Primary circuit of equipment with a
power cord to be connected to a mains
socket.
CAT.III :
Primary circuit of equipment that inputs
power directly from the distributor and the
circuit from the distributor to the mains
socket.
CAT.IV :
Circuit from the leading wire to the
distributor.
[3] NAMES OF COMPONENT UNITS
3-1 Main unit
Current Sensor (CT)
Barrier
CT trigger
Power / Function switch
HOLD button
BPF button
MAXMIN button
Display
Input Terminal (-)
Input Terminal (+)
Backlight button
3-2 Display
AUTO POWER OFF
AC
DC
Minus
polarity
Low battery indication
MAXMIN
Digit and
decimal point
BPF
DATA HOLD
UNIT
3-3 Test Lead
T
es pint s
T t )es probe(red
Finge guardr s
T t )es probe(blackPlugs
Remova tes pin covel t s r
In case of test pins cover attached: CAT.III 600V
In case of test pins cover removed: CAT.II 1000V
[4] DESCRIPTION OF FUNCTION
4-1 Power Switch and function switch
Turn this switch to turn on and off the power
and select a measuring function.
4-2 Data Hold function
When the HOLD button is pressed, the reading
indicated will be held with on the
display. The indicated reading will not change if
the input signal is changed. When this button is
pressed again, the function will be disabled and
the meter will return to the measurement mode.
Remarks:
The function will be canceled at changing
measurement functions.
4-3 BPF Button (ACmA & ACA Function)
Will switch to no-filter mode BPF 50 Hz → → BPF
60 Hz no-filter mode each time you press the
button. With this BPF function, you can extract
the frequency in which you would like to measure.
Remarks:
Frequency Response
50 Hz BPF: Using 25 Hz - decrease to 15 %
(-16 dB), 100 Hz - decrease to 4 % (-27 dB)
60 Hz BPF: Using 30 Hz - decrease to 15 %
(-16 dB), 120 Hz - decrease to 4% (-27 dB)
BPF frequency response curve
Remarks:
Readings of current are not properly in BPF
function if measuring current frequency is
other than 50/60 Hz.
If the measured current contains extremely
la r ge n oise, i t may not be pos sib l e to
measure correctly.
4-4 MAX/MIN function
When the MAX/MIN button is pressed, the
m e t e r e n t e r s M A X / M I N m o d e , f i x e s a
measuring range and turns on. Press
the button, to read MAX, MIN, current reading
sequence. Press the button for 1 sec. or more
to exit the MAX/MIN mode.
Remarks:
Function changes or functional operations will
cancel the function.
4-5 BACKLIGHT function
When the backlight button is pressed, the
backlight will be turned on. To disable the
function, the button is pressed again. The
backlight will be automatically turned off about
30 seconds after it is turned on.
4-6 AUTO POWER OFF function
The meter will go to AUTO POWER OFF function
to save battery life about 30 minutes after the
last function switch or button operation. To wake
up the meter, press any button or re-power on.
turns on while activating the function.
Remarks:
Even in the AUTO POWER OFF mode, the
tiny power is still consumed. When the meter
is not going to be used for an extended
period of time, be sure to turn off the power
switch.
To disable the function, turn the function
switch to position other than OFF while
holding the HOLD button. on the display is
turned off when the AUTO POWER OFF
function is disabled.
4-7 Low Battery indication
W h en t h e bu i l t -i n b a t t er i es h a v e b ee n
discharged and the voltage has dropped to
below about 2.2 V, appears on the display.
When the mark flickers or lights, replace both
two batteries with new ones.
[5] MEASURING PROCEDURE
1.
Do not apply an input signal exceeding
t h e m a xi m u m r a t ed i n p u t of ea c h
function.
2. During measurement, do not change
the function switch.
3. During measurement, keep your finger
behind the finger guard of the test lead
and the meter.
4. When measurement has been finished,
remove the test leads and CT from the
object measured and turn the function
switch to OFF position.
DIGITAL LEAKAGE
CLAMP METER
INSTRUCTION MANUAL
02-2205 2040 6011
DLC470
5-1 Start-up Inspection
Always conduct the start-up inspection to
ensure safety.
1.Be sure that when the meter is turned on, the
low battery indication is not flickering or
lit. If it is flickering or lit, replace the batteries
with new ones.
2.Do not use the meter if the meter or test lead
is damaged or broken.
3.Make sure the test leads are not cut.
4.
Make sure the meter and your hand are not wet.
5-2 Current Measurement ACmA
Remove the test leads from the measuring
terminals to avoid electric shock.
AC
mA
ACm curren measuremenA t t
Zero-phas currene t
clam al wirep l of s
AC
mA
ACm curren measuremenA t t
Zero-phas currene t
clam eart wirp h e
Remarks:
Clamp the conductor (cable) to measure at
the center of the trans core (CT).
To measure zero-phase current, clamp only
earth wire or all of wires like 2-core or 3-core
wire together.
The meter may malfunction in places where a
strong magnetic field is present.
Do not apply voltage and current at the same
time.
Dempa Bldg., 4-4 Sotokanda 2-Chome Chiyoda-ku, Tokyo, Japan
- MEMO - WARNING
WARNING
WARNING


Produkt Specifikationer

Mærke: Sanwa
Kategori: Måleudstyr
Model: DLC470

Har du brug for hjælp?

Hvis du har brug for hjælp til Sanwa DLC470 stil et spørgsmål nedenfor, og andre brugere vil svare dig




Måleudstyr Sanwa Manualer

Sanwa

Sanwa PDR302 Manual

29 August 2024
Sanwa

Sanwa PDR4000 Manual

29 August 2024
Sanwa

Sanwa KDP10 Manual

29 August 2024
Sanwa

Sanwa SE300 Manual

28 August 2024
Sanwa

Sanwa SE9100 Manual

28 August 2024
Sanwa

Sanwa LX20 Manual

28 August 2024
Sanwa

Sanwa LP10 Manual

28 August 2024
Sanwa

Sanwa DCL31DR Manual

28 August 2024
Sanwa

Sanwa MG5000 Manual

28 August 2024
Sanwa

Sanwa HG561H Manual

28 August 2024

Måleudstyr Manualer

Nyeste Måleudstyr Manualer

Stamos

Stamos S-LS-28 Manual

15 Oktober 2024
Stamos

Stamos S-LS-86 Manual

15 Oktober 2024
Stamos

Stamos S-LS-61 Manual

15 Oktober 2024
Stamos

Stamos S-LS-81 Manual

15 Oktober 2024
Stamos

Stamos S-LS-84 Manual

15 Oktober 2024
Stamos

Stamos S-LS-74 Manual

15 Oktober 2024
Stamos

Stamos S-LS-42 Manual

15 Oktober 2024
Stamos

Stamos S-LS-75 Manual

15 Oktober 2024
Stamos

Stamos S-LS-35 Manual

15 Oktober 2024
Stamos

Stamos S-LS-38 Manual

15 Oktober 2024