Rossmax V3 Manual
Læs nedenfor 📖 manual på dansk for Rossmax V3 (2 sider) i kategorien Medicinsk. Denne guide var nyttig for 8 personer og blev bedømt med 4.5 stjerner i gennemsnit af 2 brugere
Side 1/2
Model: V3
Suction Unit
www.rossmax.com
Rossmax Swiss GmbH,
Widnauerstrasse 1, CH-9435 Heerbrugg, Switzerland
IN3V30000000000XX
Warranty Card
This instrument is covered by a 2 year guarantee from the purchase date. The guarantee
is valid only on presentation of the guarantee card completed by the dealer confirming
purchase date or the receipt. Suction unit accessories are not included. Opening
or altering the instrument invalidates the guarantee. The guarantee does not cover
damage, accidents or noncompliance with the instruction manual. Please contact your
local seller/dealer or www.rossmax.com.
Customer Name: ____________________________________________________
Address: ___________________________________________________________
Telephone: _________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store where purchased:
____________________________________________________________________
Float (12a) PP O O X
Connecting tubing
(Ø1/4”, 180 cm) (16) PVC O X X
Connecting tubing
(Ø8 mm, 14 cm) (14) Silicone O O X
O: Applicable, X: Not applicable
Note:
1. The collection canister and tubing after disinfecting should be dated and stored
in clean and dry areas to avoid recontamination.
2. Do not use any cleaners or disinfectants that contain ammonia, benzene and/or
acetone to clean the unit.
3. Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
4. Dispose of the suction catheter in accordance with the local laws and regulations.
Collection canister – 1400 ML
Filter
Silicone and PVC tubing
Maintenance
Rules for returning and repairing
Microbial contamination
Accessories (optional)
Technical Specications
Troubleshooting
Electromagnetic Compatibility Information
Symbols
The collection canister consists of an overow valve, a bottle in PC (polycarbonate)
material and must be replaced if there are visible signs of the damage.
Note:
1. Keep the device vertical to allow the overow valve to function correctly.
2. Do not overturn the collection canister during use, in order to prevent the
intervention of the overow valve; should this occur, switch the aspirator o
and detach the tubing connected to the antibacterial lter.
3. To accelerate aspiration and simplify cleaning, it is recommended to put water
in the collection canister before using.
4. Suction at the patient is automatically obstructed when the liquid level reaches
the overow valve located on the underside of the canister lid.
5. Contents of the container should be emptied when automatic overow valve is
activated. Further suctioning could cause damage to the vacuum pump.
The antibacterial lter supplied with Rossmax aspirators must be strictly replaced
for each new patient, or if overow occurs or the lter is saturated. Should
the aspirator be used on the same patient, the lter for single-patient use is
recommended to be replaced after every application. It should also be replaced if
the unit is used with a patient whose risk of contamination is unknown. The lter
cannot be sanitized, disinfected, or sterilized.
Note:
1. Do not use the aspirator without the antibacterial lter, the unit could be
damaged. From a bacteriological point of view, it becomes dangerous for the
patient.
2. Do not substitute any other material for this antibacterial lter. Substitution may
lead to contamination or poor performance; use only Rossmax lters.
The tubing is recommended to be single use, but the number of cycles of
disinfection and/or cleaning is strictly linked to the employment of the said tubing.
Therefore, after each cleaning cycle, it is up to the nal user to verify whether the
tubing is suitable for reuse. The component must be replaced if there are visible
signs of the decay of the tubing.
The suction equipment does not need maintenance or lubrication. It is, however
to inspect the unit and accessories before each use.
DANGER:
Electric shock hazard. Do not attempt to open or remove cabinet, there are no
user-serviceable internal components. If service is required, return the device to
your local distributor or authorized Rossmax service center. Opening or tampering
with the unit will void warranty.
Rossmax requires the following procedures to be carried out to protect
the instrument and the safety of all who come in contact with it.
Before returning an instrument for repair, the external surfaces and all accessories
MUST be carefully disinfected with a cloth soaked in methylated spirits or
hypochlorite-based solution. The instrument and accessories should then
be placed in a bag with a note outlining the disinfection undertaken. Failure
to follow this procedure will result in the instrument being returned to the
purchaser unrepaired. Instruments returned for repair MUST be accompanied
by a description of the problem. Rossmax will not be responsible for damage
caused through improper use. To avoid such damage, please read the instruction
carefully. Rossmax will not be responsible for damage of accessories. Accessories
will be replaced at the customer’s expense
In the presence of pathologies with the risk of infection and microbial contamination,
we recommend personal use of the accessories, collection canister, connecting
tubing and antibacterial lter and all accessories should be thoroughly cleaned
and disinfected after each use.
Note: Use only with original Rossmax accessories.
Name Model No.
Connecting PVC tubing (Ø1/4”-length
180cm) SU_AC_001_00
Collection canister (1400ml) SU_AC_003_00
Antibacterial lter kit (Antibacterial lter +
2 Silicone tubing Ø8mm-length 14cm) SU_AC_004_00
Antibacterial lter SU_AC_005_00
Connecting Silicone tubing
(Ø8mm-length 14cm) SU_AC_006_00
AC/DC switching Adaptor (12V/2A)
(EU plug) SU_AD_001_00
(MODEL No.: UES24 LCP-120200SPA)
AC/DC switching Adaptor (12V/2A)
(US plug) SU_AD_002_00
(MODEL No.: UES24 LCP-120200SPA)
AC/DC switching Adaptor (12V/2A)
(UK plug) SU_AD_003_00
(MODEL No.: UES24 LCP-120200SPA)
Suction class: High Vacuum / Low Flow
Adaptor: Input: 100-240V, 50/60Hz; Output: DC 12V 2.0A
Max. vacuum @ pump inlet: -73 kPa/ -550 mmHg (mmHg for reference only)
Max air ow @ pump inlet: 16L/min
Power Consumption: ≤ 24W
Collection canister: 1,400 ml
Weight: Approx. 1.92 (kgs)
Machine dimension (with-
out collection canister): 363 (W) × 175 (D) × 254.5 (H) mm (Incl. rubber pad)
Operating conditions: Temperature: min: 10°C; max. 40°C;
Relative humidity: ≤90% non-condensing
Atmospheric pressure: Min. 700 hPa; max. 1060 hPa
Storage and transportation
conditions:
Temperature min: -20°C; max. 60°C;
Relative humidity: ≤90% non-condensing
Atmospheric pressure: Min. 700 hPa; max. 1060 hPa
IP Classication: IP21: Protection against harmful ingress of water
and particulate matter
If, after checking the described conditions below the device is still not working
properly, we recommend that you consult your local distributor or authorized
Rossmax service center.
Problem Possible cause Solution
The device does not
work
The AC adaptor has not
been correctly inserted
into the socket of the
device or the power
socket
Correctly insert the AC
adaptor into the sockets
Power switch has not
been activated.
Check if the power switch
is turned to the “I” (ON)
position
Lack of suction
The collection canister
lid has not been correctly
positioned on the canister
Correctly position the
collection canister lid
Leaks or cracks in canister
assembly Replace the canister
Filter blocked Replace the filter
Blocked float Incrustation on the float
Remove the canister
lid and overflow valve
and extract the float.
Continue by carrying out
the cleaning procedures
as described in the
paragraph “Sanitization,
Disinfection operations”
Vacuum power poor
and/or non-existent
Vacuum flow adjustment
knob/ vacuum regulator
completely open
Fully close the adjustment
knob and check the
vacuum power
Filter blocked Replace the filter
Connecting tubing to
the device or collection
canister is clogged, bent,
disconnected or broken;
or suction catheter is
twisted or blocked
Check the condition
of the tubing, replace
it if blocked or broken
and correctly connect it
as per the Assembling
Procedures of this manual
Canister lid overflow valve
closed or blocked
Unblock the overflow
valve, keep the device
upright
Dirty, blocked or
damaged pump
Take the device to
your local distributor
or authorized Rossmax
service center
Class II device
Type BF applied part
Attention: check the instructions before using this device
+More vacuum
SN Device serial number
Risk: Electrocution.
Consequence: Death.
Do not use the device while taking a bath or a shower
ΙSwitch on
ΟSwitch o
Direct current
1639 CE Marking medical ref. Dir 93/42/EEC amended by 2007/47/EC
Date of manufacture
Manufacturer
Attention: Consult accompanying documents
WARNING: The symbol on this product means that it’s an electronic
product and following the European directive 2012/19/EU the elec-
tronic products have to be disposed on your local recycling centre for
safe treatment.
OBM_IB_SU-V3_
EN_SW_ver2138
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the V3,
including cables specied by the manufacturer. Otherwise, degradation of the performance of this
device could result.
If higher IMMUNITY TEST LEVELS than those specied in Table 9 are used, the minimum separation
distance may be lowered. Lower minimum separation distances shall be calculated using the
equation specied in 8.10.
Manufacturer’s declaration-electromagnetic immunity
The V3 is intended for use in the electromagnetic environment specied below. The customer or
the user of the V3 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the V3 including cables, than the
recommended separation distance calcu-
lated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz, d
= 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rat-
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d
is the recommended separation distance in
metres (m).
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by
absorption and reection from structures, objects and people.
More information on EMC compliance of the device can be obtained from Rossmax website: www.
rossmax.com.
Produkt Specifikationer
Mærke: | Rossmax |
Kategori: | Medicinsk |
Model: | V3 |
Har du brug for hjælp?
Hvis du har brug for hjælp til Rossmax V3 stil et spørgsmål nedenfor, og andre brugere vil svare dig
Medicinsk Rossmax Manualer
23 August 2024
21 August 2024
19 August 2024
19 August 2024
16 August 2024
12 August 2024
12 August 2024
11 August 2024
11 August 2024
8 August 2024
Medicinsk Manualer
- Medicinsk GIMA
- Medicinsk Medtronic
- Medicinsk Vermeiren
- Medicinsk Pulsar
- Medicinsk Dittmann
- Medicinsk Hybernite
- Medicinsk Diley Dreams
- Medicinsk VQ OrthoCare
- Medicinsk Rhino Horn
Nyeste Medicinsk Manualer
27 August 2024
25 August 2024
25 August 2024
25 August 2024
25 August 2024
25 August 2024
25 August 2024
24 August 2024
23 August 2024
22 August 2024